Lars Kasper [1,2,†], Maria Engel [1], Jakob Heinzle [2], Matthias Mueller-Schrader [2], Nadine N. Graedel [1,3], Jonas Reber [1], Thomas Schmid [1], Christoph Barmet [1], Bertram J. Wilm [1], Klaas Enno Stephan [2,4,5], Klaas P. Pruessmann [1]
-
Institute for Biomedical Engineering, ETH Zurich and University of Zurich, Gloriastrasse 35, 8092 Zurich, Switzerland
-
Translational Neuromodeling Unit, Institute for Biomedical Engineering, University of Zurich and ETH Zurich, Wilfriedstrasse 6, 8032 Zurich, Switzerland
-
Wellcome Centre for Integrative Neuroimaging, FMRIB, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, United Kingdom
-
Wellcome Centre for Human Neuroimaging, University College London, London WC1N 3BG, UK
-
Max Planck Institute for Metabolism Research, 50931 Cologne, Germany
†Current Address: Techna Institute, University Health Network Toronto, Canada
-
This work reports the first fMRI study at 7T with high-resolution spiral readout gradient waveforms.
-
We achieve spiral fMRI with sub-millimeter resolution (0.8 mm, FOV 230 mm), acquired in a single shot (36 slices in 3.3 s).
-
Spiral images exhibit intrinsic geometric congruency to anatomical scans, and spatially specific activation patterns.
-
Image reconstruction rests on a signal model expanded by measured trajectories and static field maps, inverted by cg-SENSE.
-
We assess generalizability of the approach for spiral in/out readouts, providing two images per shot (1.5 mm resolution).
Spiral fMRI has been put forward as a viable alternative to rectilinear echo-planar imaging, in particular due to its enhanced average k-space speed and thus high acquisition efficiency. This renders spirals attractive for contemporary fMRI applications that require high spatiotemporal resolution, such as laminar or columnar fMRI. However, in practice, spiral fMRI is typically hampered by its reduced robustness and ensuing blurring artifacts, which arise from imperfections in both static and dynamic magnetic fields.
Recently, these limitations have been overcome by the concerted application of an expanded signal model that accounts for such field imperfections, and its inversion by iterative image reconstruction. In the challenging ultra-high field environment of 7 Tesla, where field inhomogeneity effects are aggravated, both multi-shot and single-shot 2D spiral imaging at sub-millimeter resolution was demonstrated with high depiction quality and anatomical congruency.
In this work, we further these advances towards a time series application of spiral readouts, namely, single-shot spiral BOLD fMRI at 0.8 mm in-plane resolution. We demonstrate that high-resolution spiral fMRI at 7 T is not only feasible, but delivers both excellent image quality, BOLD sensitivity, and spatial specificity of the activation maps, with little artifactual blurring. Furthermore, we show the versatility of the approach with a combined in/out spiral readout at a more typical resolution (1.5 mm), where the high acquisition efficiency allows to acquire two images per shot for improved sensitivity by echo combination.
Functional MRI (fMRI) is presently the most prominent technique to study human brain function non-invasively, owing to its favorable spatiotemporal resolution regime with appealing functional sensitivity. Within this regime, specific research questions require different trade-offs between spatial and temporal resolution. On the one hand, ultra-high spatial resolution fMRI (with sub-millimeter voxel size) successfully targets smaller organizational structures of the brain, such as cortical laminae (Fracasso et al., 2016; Huber et al., 2017a; Kashyap et al., 2018; Kok et al., 2016; Lawrence et al., 2018; Martino et al., 2015; Muckli et al., 2015; Siero et al., 2011) and columns (Cheng et al., 2001; Feinberg et al., 2018; Yacoub et al., 2008). For subcortical sites, due to the limited signal-to-noise ratio (SNR), high-resolution (1-1.5 mm) fMRI is more prevalent (but see (Wang et al., 2020)) to characterize, for example, the superior (Savjani et al., 2018; Singh et al., 2018) and inferior colliculi (De Martino et al., 2013; Sitek et al., 2019), as well as the subthalamic nucleus (de Hollander et al., 2017) and midbrain (D’Ardenne et al., 2008). However, both high and ultra-high spatial resolution fMRI require compromises on field of view (FOV) coverage or temporal bandwidth, i.e., volume repetition time (TR). On the other hand, fast sequences with TRs on the order of 0.5 seconds and below are important for advanced analysis approaches, for example, to adequately sample physiological fluctuations (Lewis et al., 2016; Smith et al., 2013; Uğurbil et al., 2013), at the expense of lowering spatial resolution (2–4 mm).
One means to simultaneously advance the spatial and temporal resolution boundaries of fMRI is to maximize acquisition efficiency, i.e., sampled k-space area (or volume) per unit time. Therefore, fMRI nowadays almost exclusively relies on rectilinear echo-planar imaging (EPI, (Cohen and Schmitt, 2012; Mansfield, 1977; Schmitt et al., 2012)), where acquisition efficiency is favorable due to optimal acceleration and high terminal velocity along the straight k-space lines traversed.
To expand spatiotemporal resolution beyond the capabilities of EPI alone, the main strategy has been parallel imaging acceleration (Griswold et al., 2002; Pruessmann et al., 1999; Sodickson and Manning, 1997), in combination with simultaneous multi-slice or 3D excitation (Breuer et al., 2006; Larkman et al., 2001; Poser et al., 2010; Setsompop et al., 2012). In terms of k-space coverage per unit time, the benefit of parallel imaging lies in expanding the cross section of the k-space neighborhood covered along the readout trajectory (Pruessmann, 2006), i.e., a band in 2D or tube in 3D.
However, another key determinant of acquisition efficiency or speed of coverage is average velocity along the trajectory, i.e., instantaneous gradient strength. On this count, EPI is wasteful because it includes many sharp turns traversed at low speed due to the limited gradient slew rate.
Substantially higher average k-space speed and thus acquisition efficiency for fMRI is achieved with spiral trajectories (Barth et al., 1999; Glover, 2012; Noll et al., 1995), which avoid sharp turns by distributing curvature more evenly (Ahn et al., 1986; Likes, 1981; Meyer et al., 1992). Typically, single-shot variants winding out of k-space center, e.g., on an Archimedean spiral, are prevalent (Glover, 1999; Meyer et al., 1992; Weiger et al., 2002), but different acquisition schemes, such as spiral-in (Börnert et al., 2000) or combined in/out readouts (Glover and Law, 2001; Glover and Thomason, 2004) have been proposed. High-resolution fMRI studies have occasionally employed spirals as well (Jung et al., 2013; Singh et al., 2018), including first applications of laminar fMRI (Ress et al., 2007) and regionally optimized acquisitions, e.g., for the hippocampus (Preston et al., 2010) or superior colliculus (Katyal et al., 2010; Savjani et al., 2018). Common to these approaches is a reduction of acquisition efficiency in favor of robustness by acquiring k-space in multiple shots with shorter spiral readouts.
Despite these efforts, routine use of spiral fMRI has not been established, due to the following three challenges (Block and Frahm, 2005; Börnert et al., 1999): First, spirals are sensitive to imperfect magnetic field dynamics (drifts, eddy currents and other gradient imperfections) which lead to blurring and image distortions. Secondly, non-uniformity of the static B0 field, caused by varying susceptibility of the imaged tissues, likewise causes blurring (Bernstein et al., 2004, Chap. 17). Finally, in combination with parallel imaging, spirals pose a somewhat greater reconstruction challenge than Cartesian trajectories (Pruessmann et al., 2001).
Recently, these obstacles have been overcome (Engel et al., 2018; Kasper et al., 2018; Wilm et al., 2017) by (1) employing an expanded signal model that incorporates coil sensitivity encoding as well as independently measured static and dynamic field imperfections (Wilm et al., 2011), and (2) the inversion of this model by an accompanying iterative image reconstruction (Barmet et al., 2005; Man et al., 1997; Pruessmann et al., 2001; Sutton et al., 2003). This approach enabled the use of long spiral readouts (on the order of 50 ms at 7 Tesla), while maintaining high image quality and anatomical fidelity. In particular, such enhanced spiral acquisition efficiency was demonstrated by accomplishing T2*-weighted images with a nominal in-plane resolution of 0.8 mm in a single shot. Ultimately, these findings hold promise that spiral fMRI can now indeed profit from the theoretical benefits of enhanced acquisition efficiency to expand the spatiotemporal boundaries of fMRI.
Based on these advances in expanded signal modeling and inversion, in this work, we explore the feasibility and utility of sub-millimeter single-shot spiral fMRI. Specifically, we first assess image quality and temporal stability of fMRI time series obtained with the expanded signal model and algebraic reconstruction. We further evaluate the resulting functional sensitivity and spatial specificity of reference activation patterns, elicited by an established visual quarter-field stimulation paradigm.
Finally, we explore the versatility of the approach with a combined in/out spiral readout at a more typical resolution (1.5 mm). Here, two images per shot can be acquired, translating the high acquisition efficiency of the spiral into enhanced functional sensitivity by echo combination (Glover and Law, 2001; Glover and Thomason, 2004; Law and Glover, 2009).
All data was acquired on a Philips Achieva 7 Tesla MR System (Philips Healthcare, Best, The Netherlands), with a quadrature transmit coil and 32-channel head receive array (Nova Medical, Wilmington, MA, USA).
Concurrent magnetic field monitoring was performed using 16 fluorine-based NMR
field probes, which were integrated into the head setup via a laser-sintered
nylon frame positioned between transmit and receive coil (Fig. 1 in Engel et
al., 2018). Probe data were recorded and preprocessed (filtering, demodulation)
on a dedicated acquisition system (Dietrich et al., 2016a). The final extraction
of probe phase evolution and projection onto a spherical harmonic basis set
(Barmet et al., 2008) was performed on a PC, yielding readout time courses of
global phase
For the fMRI experiments, visual stimulus presentation utilized VisuaStim LCD goggles (Resonance Technology Inc., Northridge, CA, USA). A vendor-specific respiratory bellows and finger pulse plethysmograph recorded subject physiology, i.e., respiratory and cardiac cycle.
Seven healthy subjects (4 female, mean age 25.7 +/- 4.1 y) took part in this study, after written informed consent and with approval of the local ethics committee. One subject was excluded from further analysis due to reduced signal in multiple channels of the head receive array. Thus, six subjects were analyzed for this study.
The paradigm, a modified version of the one used in (Kasper et al., 2014), comprised two blocks of 15 s duration that presented flickering checkerboard wedges in complementary pairs of the visual quarter-fields. In one block, upper left and lower right visual field were stimulated simultaneously (condition ULLR), while the other block presented the wedges in the upper right and lower left quarter-fields (condition URLL). These stimulation blocks were interleaved with equally long fixation periods. To keep subjects engaged, they had to respond to slight contrast changes in the central fixation cross via button presses of the right hand. A single run of the paradigm took 5 min (5 repetitions of the ULLR-Fixation-URLL-Fixation sequence). For both of the spiral sequence designs (high-resolution spiral-out and combined spiral in/out, see next section), a single run of the paradigm was performed per subject.
Spiral fMRI was based on a slice-selective multi-slice 2D gradient echo sequence (Fig. 1) with custom-designed spiral readout gradient waveforms. For every third slice, i.e., a TR of 270 ms, concurrent field recordings were performed on the dedicated acquisition system (Dietrich et al., 2016a), with NMR field probes being excited a few milliseconds prior to readout gradient onset (Fig. 1, bottom row, (Engel et al., 2018)).
For the spiral trajectories, we selected two variants that had previously provided high-quality images in individual frames (Fig. 2 in (Engel et al., 2018)): a high-resolution case winding out of k-space center on an Archimedean spiral (spiral-out, Fig. 1, black gradient waveform), and a combined dual-image readout first spiraling into k-space center, immediately followed by a point-symmetric outward spiral (spiral in/out (Glover and Law, 2001)), Fig. 1, blue gradient waveform).
The spiral-out gradient waveform was designed to deliver the highest spatial resolution possible under several constraints. First, targeting maximum acquisition efficiency in 2D commands a single-shot 2D readout, because the sequence overhead, i.e., time spent without sampling k-space, accrues for each new excitation. Second, the parallel imaging capability of our receiver array at 7 T allowed for an in-plane acceleration factor of R = 4 (determining the spacing of the spiral revolutions, i.e., FOV). We based this choice on previous experience with spirals of such undersampling using this setup (Engel et al., 2018; Kasper et al., 2018), which were free of aliasing artifacts or prohibitive g-factor noise amplification. Third, the requirement of concurrent field recordings for the whole spiral readout limited its maximum duration to below 60 ms. This is the approximate lifetime of the NMR field probe signal, after which complete dephasing occurs in a subset of probes for this specific setup, governed by their T2* decay time of 24 ms (Engel et al., 2018) and distance from iso-center when applying higher-order shims. Finally, the gradient system specifications constrain the maximum possible resolution (or k-space excursion) of an Archimedean spiral with prescribed FOV and duration. Here, we used the optimal control algorithm by (Lustig et al., 2008) to design time-optimal spiral gradient waveforms of 31 mT/m maximum available gradient amplitude, and a 160 mT/m/ms slew rate limit, chosen for reduced peripheral nerve stimulation.
Overall, these requirements led to a spiral-out trajectory with a nominal in-plane resolution of 0.8 mm (for a FOV of 230 mm), at a total readout acquisition time (TAQ) of 57 ms. BOLD-weighting was accomplished by shifting the readout start, i.e., TE, to 20 ms.
For the spiral in/out, we followed the same design principles, targeting a minimum dead time after excitation, and a symmetric readout centered on a TE of 25 ms, slightly shorter than reported T2* values in cortex at 7 T (Peters et al., 2007). This resulted in a gradient waveform lasting 39 ms, with a nominal resolution of 1.5 mm for each half-shot of the trajectory.
All other parameters of both spiral sequences were shared, in order to facilitate comparison of their functional sensitivity. In particular, slice thickness (0.9 mm) and gap (0.1 mm) were selected for near-isotropic sub-mm resolution for the spiral-out case, while still covering most of visual cortex. For each slice, the imaging part of the sequence (Fig. 1) was preceded by a fat suppression module utilizing Spectral Presaturation with Inversion Recovery (SPIR, (Kaldoudi et al., 1993)).
The sequence duration totaled 90 ms per slice for the spiral-out sequence (TE 20 ms + TAQ 60 ms + SPIR 10 ms), which was maintained for the spiral in/out, even though a shorter imaging module would have been possible. To arrive at a typical volume repetition time for fMRI, we chose to acquire 36 slices (TR 3.3 s). Each functional run comprised 100 volume repetitions, amounting to a scan duration of 5.5 min.
Image reconstruction rested on an expanded model of the coil signal
s_{\gamma}\left( t \right) = \int_{V}^{}{c_{\gamma}\left( \mathbf{r} \right) \cdot m\left( \mathbf{r} \right) \cdot e^{i\sum_{l}^{}{k_{l}\left( t \right)b_{l}\left( \mathbf{r} \right)}} \cdot e^{i\Delta\omega_{0}(\mathbf{r})t} \cdot \text{dV}} |
---|
with coil index
For 2D spiral imaging without strong higher order eddy currents (e.g., as
induced by diffusion encoding gradients), this model can be computationally
reduced (Engel et al., 2018) to facilitate iterative inversion. To this end, we
(1) considered only field dynamics contributing to global phase
s_{\gamma}\left( t \right) = \int_{V}^{}{c_{\gamma}\left( \mathbf{r} \right) \cdot m\left( \mathbf{r} \right) \cdot e^{i\left( k_{0}\left( t \right) + \mathbf{k}\left( t \right) \cdot \mathbf{r} \right)} \cdot e^{i\Delta\omega_{0}\left( \mathbf{r} \right)t} \cdot \text{dV} = \ e^{i\left( k_{0}\left( t \right) + \mathbf{k}\left( t \right) \cdot \mathbf{r}{\mathbf{0}} \right)}\int{V}^{}{c_{\gamma}\left( \mathbf{r} \right) \cdot m\left( \mathbf{r} \right) \cdot e^{i\mathbf{k}\left( t \right) \cdot \left( \mathbf{r}\mathbf{-}\mathbf{r}{\mathbf{0}} \right)} \cdot e^{i\Delta\omega{0}\left( \mathbf{r} \right)t} \cdot \text{dV}}} |
---|
For the demodulated coil signal
{\tilde{s}}{\left( \gamma,\tau \right)} = \sum{\rho}^{}{E_{\left( \gamma,\tau \right),\rho}m_{\rho}} |
---|
with sampling time index
The matrix-vector form of eq. (3) is a general linear model,
\tilde{\mathbf{s}} = E\ \mathbf{m} |
---|
and can be efficiently solved iteratively by a conjugate gradient (CG) algorithm (Pruessmann et al., 2001; Shewchuk, 1994). As mentioned above, the restriction to first order field dynamics enables acceleration of the ensuing matrix-vector multiplications by (reverse) gridding and fast Fourier transform (FFT) (Beatty et al., 2005; Jackson et al., 1991). Off-resonance effects are also efficiently approximated by FFT using multi-frequency interpolation (Man et al., 1997).
This image reconstruction algorithm was applied equivalently to the spiral-out and spiral in/out data with a fixed number of 10 iterations and no further regularization (e.g., Tikhonov). Note, however, that for the latter both field recordings and coil data were split into their in- and out-part and reconstructed separately, yielding two images per shot.
Taken together, the in-house Matlab R2018a (The MathWorks, Natick, MA, USA) implementation of this algorithm led to total reconstruction times on a single CPU core of about 10 min per slice for the high-resolution spiral-out image and 1.5 minutes for the spiral-in or -out image. In order to reconstruct the 3600 2D images per fMRI run, reconstruction was parallelized over slices on the university’s CPU cluster. Depending on cluster load, reconstructions typically finished over night for the high-resolution spiral out, and within 2 h for the spiral in/out data.
The auxiliary input data for the expanded signal model, i.e., spatial maps for
static B0 field inhomogeneity
Sensitivity maps were then computed from the first-echo image, normalizing single coil images by the root sum of squares over all channels, while the B0 map was calculated by regressing the pixel-wise phase evolution over echo images. Both maps were spatially smoothed and slightly extrapolated via a variational approach (Keeling and Bammer, 2004).
The suitability of the raw imaging data for high-resolution fMRI was assessed in terms of both sensitivity and spatial specificity. No smoothing was performed for any analysis in this section.
For sensitivity, we evaluated the temporal statistics of the images, i.e., signal-to-fluctuation noise ratio (SFNR), standard deviation (SD) and coefficient of variation (CoV) maps (Welvaert and Rosseel, 2013), defined as
\text{CoV} = \frac{1}{\text{SFNR}}\ , |
---|
where the bar denotes averaging over volumes of a run.
Our assessment of spatial specificity was based on the ME reference scan, which
exhibits a high geometric veracity due to its spin-warp nature, i.e., high
bandwidth. We overlaid the contour edges (intensity isolines) of the mean (over
echoes) of the ME images onto the mean spiral images
To reduce the impact of subject motion on both assessments, the volumes of the fMRI time series were first realigned to each other using a 6-parameter rigid-body within-contrast registration, as implemented in SPM (Friston et al., 1996). Then, the mean ME scan was co-registered to the resulting mean realigned fMRI scan, Importantly, both operations were limited to six-parameter rigid-body registration, such that nonlinear geometric distortions between sequences were not corrected through this preprocessing step. Furthermore, to facilitate visual comparison and contour edge creation, mean ME and spiral images were bias-field corrected using unified segmentation (Ashburner and Friston, 2005).
Furthermore, for quantitative assessment, we extracted contour lines from the
thresholded gray matter tissue probability maps (
All computations were performed in Matlab R2019b, using the Unified NeuroImaging Quality Control Toolbox (UniQC, (Bollmann et al., 2018; Frässle et al., 2021)), and SPM12 (Wellcome Centre for Human Neuroimaging, London, UK, http://www.fil.ion.ucl.ac.uk/spm/).
The main goal of this analysis was to assess the functional sensitivity and spatial specificity of the spiral fMRI sequences at the single-subject level under standard paradigm and preprocessing choices. Note that, unless explicitly stated otherwise, all activation maps and their quantification (e.g., cluster extent, peak t-values) are therefore reported after smoothing.
Equivalent preprocessing steps were applied to all spiral fMRI runs using SPM12. After slice-timing correction, we employed the pipeline described in the previous section (realignment, co-registration, bias-field correction via unified segmentation). Finally, the functional images were slightly smoothed with a Gaussian kernel of 0.8mm FWHM, i.e., the voxel size of the high-resolution scan.
The general linear model (GLM) contained regressors of the two stimulation blocks (ULLR and URLL) convolved with the hemodynamic response function (HRF), as well as nuisance regressors for motion (6 rigid-body parameters) and physiological noise (18 RETROICOR (Glover et al., 2000) regressors, as specified in (Harvey et al., 2008)), extracted by the PhysIO Toolbox (Kasper et al., 2017).
To characterize functional sensitivity, we evaluated the differential t-contrasts +/- (ULLR-URLL) and report results at an individual voxel-level threshold of p < 0.001 (t > 3.22). For quantification in the results tables, we report activations under whole-brain family-wise error (FWE) correction at the cluster level (p < 0.05), given this voxel-level threshold (p < 0.001) for cluster definition. If not noted otherwise in the figure, we omitted multiple-comparison correction in the visualization of the activation maps, to study their spatial extent and specificity.
Spatial specificity of the activation was qualitatively assessed by overlaying the thresholded t-contrast maps for both contrasts onto the anatomically veridical mean ME image. We checked whether activation patterns were restricted to gray matter regions of visual cortex, as well as whether the spatial separation and symmetry of activations were linked to distinct quarter-field stimulation patterns, as expected by the retinotopic organization of visual cortex (Engel et al., 1997; Wandell et al., 2007; Warnking et al., 2002). Furthermore, we evaluated the individual contrasts for the ULLR and URLL stimulation blocks to assess the spatial overlap of their activation patterns as an alternative measure of functional specificity (since the differential contrasts cannot overlap by design).
The quantification of functional spatial specificity relied on the tissue probability maps extracted via unified segmentation from the structural scan (mean ME). To reduce the number of uncategorized voxels, we chose a liberal exceedance threshold of 60 % to define the individual tissue classes (gray matter (GM), white matter (WM), cerebrospinal fluid (CSF)). Of the remaining voxels, those that exceeded 30 % probability for two tissue classes were categorized as gray/white matter boundary (GM/WM interface) or pial surface (GM/CSF interface). All other voxels were labeled as ambiguous. We then evaluated the share of significantly activated voxels for the differential t-contrasts +/- (ULLR-URLL) after multiple comparison correction (p < 0.05 cluster-level FWE corrected with a cluster-forming voxel threshold of p < 0.001). This analysis was performed for both the whole imaging FOV and within the ROI of early visual cortex, as defined at the end of the previous section (2.5.1). We repeated this analysis for activation maps derived from unsmoothed data to study the impact of smoothing on spatial specificity.
This overall analysis procedure was performed for the spiral-out as well as the individual spiral-in and spiral-out image time series reconstructed from the spiral in/out data. As spiral in/out sequences are predominantly selected for their potential gain in functional sensitivity when combining spiral-in and spiral-out images (Glover and Law, 2001), we additionally repeated the BOLD fMRI analysis for such a surrogate dataset (“in/out combined”), but omitted the quantitative analysis of spatial specificity. We chose a signal-weighted combination per voxel (Glover and Thomason, 2004), which is considered the most practical approach for echo combination (Glover, 2012):
w = \frac{\overline{m_{1}}}{\overline{m_{1}} + \overline{m_{2}}}\ , |
---|
with
Image reconstruction was performed by an in-house custom Matlab implementation of the cg-SENSE algorithm (Pruessmann et al., 2001). A demonstration of that algorithm is publicly available on GitHub (https://github.com/mrtm-zurich/rrsg-arbitrary-sense), with a static compute capsule for reproducible online re-execution on CodeOcean (Patzig et al., 2019), which were created in the context of the ISMRM reproducible research study group challenge (Maier et al., 2021; Stikov et al., 2019), albeit without the multi-frequency interpolation employed here. An example reconstruction pipeline including static B0 correction for the spiral data presented here is available on GitHub as well (https://github.com/mrikasper/julia-recon-advances-in-spiral-fmri), utilizing MRIReco.jl (Knopp and Grosser, 2021)*, an MRI reconstruction framework written in Julia* (Bezanson et al., 2017)*.*
Image and fMRI analyses were performed using SPM12 (https://www.fil.ion.ucl.ac.uk/spm, distributed under GPLv2) and the in-house developed UniQC Toolbox (Bollmann et al., 2018), publicly available under a GPLv3 license as a beta release within the TAPAS software collection (https://www.translationalneuromodeling.org/tapas, (Frässle et al., 2021)).
All custom analysis and data visualization scripts for this publication are
available on https://github.com/mrikasper/paper-advances-in-spiral-fmri. This
includes both a one-click analysis (main.m
) to rerun all image statistics and
fMRI analyses, as well as the automatic re-creation of all figure components
found in this manuscript (main_create_figures.m
), utilizing the UniQC Toolbox.
More details on installation and execution of the code can be found in the
README.md
file in the main folder of the repository.
Data from this study is publicly available as part of the ETH Research
Collection (doi:10.3929/ethz-b-000487412, (Kasper et al., 2021)) and described
in more detail in the accompanying Data in Brief Article, according to the FAIR
(Findable, Accessible, Interoperable, and Re-usable) data principles (Wilkinson
et al., 2016). For one subject (SPIFI_0007
), this includes both the
reconstructed images in NIfTI format with behavioral and physiological logfiles,
to validate the analysis scripts, as well as raw coil and trajectory data in
ISMRMRD format (in the patient coordinate system), together with the
corresponding B0 and sensitivity maps (NIfTI).
For the other datasets, we did not obtain explicit subject consent to share all raw data in the public domain. However, we do provide the magnetic field evolution time series as ISMRMRD files (in the scanner coordinate system) within the ETH Research Collection. Mean spiral fMRI images with corresponding activation t-maps for all subjects are also made available on NeuroVault for interactive viewing ((Gorgolewski et al., 2015), https://neurovault.org/collections/6086/).
Finally, for further data dissemination, montage views of the presented image quality metrics and statistical map overlays containing all slices and subjects are included in the supplementary materials (high-resolution spiral-out: SM 1, spiral in/out: SM 2).
In the following, we mainly present images from individual subjects (S7: Figs. 2,3, S2: Figs. 4,6,9, S3: Fig. 5). However, as illustrated by Fig. 7, as well as supplementary materials SM 1 and SM 2, results were comparable for all six analyzed datasets.
The mean images (one run of subject S7, after realignment) of the high-resolution spiral-out sequence exhibit good image quality, rich in T2* contrast and anatomical detail (Fig. 2A). In the center of the brain, no blurring is apparent, and anatomical boundaries can be clearly delineated, e.g, the optic radiation, down to the single-voxel extent. Moderate residual imaging artifacts (local ringing, blurring) are visible in the orbitofrontal areas, at some brain/skull boundaries, and in the vicinity of larger muscles and fat deposits, e.g., the temporal muscles. For more inferior slices, signal dropouts can be identified at typical sites of through-plane dephasing, e.g., in the temporal lobe above the ear canals (SM 1, subject 6) or in the orbitofrontal cortex (SM 1, subject 5). Individual frames of the time series exhibit similar features (Fig. 2B), though somewhat noisier, as expected because of the reduced SNR.
Interestingly, the mean of the corresponding raw phase images also contains high anatomical detail and few phase wraps (Fig. 2C), which are again located at the interface between brain and skull or close to air cavities. Note that the unwrapped appearance of the phase image is a feature of the B0-map based correction (Kasper et al., 2018) and does not require any postprocessing.
Mapping the temporal statistics of the spiral image time series (Fig. 3, Table
- proves its sufficient stability for functional imaging in all slices. The SFNR images (Fig. 3A) are rather homogeneous, with mean values of 15.3 +/- 1.1 in cortical gray matter, averaged over subjects (Table 1). A slight reduction for central brain regions is visible due to the diminished net sensitivity of the receiver array. Notably, no structured noise amplification through bad conditioning of the undersampled reconstruction problem (g-factor penalty) is discernible in this area.
The SD images (Fig. 3B) corroborate this impression, showing peak values in CSF-filled (lateral ventricles) and highly vascularized areas (insula, ACC). These noise clusters presumably stem from fluctuations through cardiac pulsation and are not specific to spiral acquisitions. However, for the raised SD values in voxels close to the cortex borders, it is unclear whether also CSF fluctuations, the BOLD effect itself, or rather time-varying blurring due to unaccounted magnetic field fluctuations contribute. This is scrutinized in the GLM analysis below. Additionally, for the CoV images (Fig. 3C), the internal capsule appears prominently, presumably due to its reduced average signal level.
In terms of spatial specificity, overlaying contour edges of the anatomical reference (mean ME spin-warp image, subject S2) (Fig. 4A) onto the mean spiral-out image suggests a geometrically very faithful depiction of the anatomical interfaces (Fig. 4B,F). Boundaries of the ventricles and gray to white matter are congruent in general, also for the visual cortex relevant to the later fMRI analyses. Some regions of the spiral-out images suffer from ringing (yellow arrow) or signal dropout (white arrow), most likely due to through-plane dephasing and incomplete correction of in-plane B0 inhomogeneity( Fig. 4E).
Incorporating the mean images of the spiral in/out sequence into the comparison confirms the nature of these artifacts (Fig. 4C,D,G,H). The in-part images (Fig. 4C) are devoid of these artifacts and match the anatomical reference almost completely in terms of edge contours (Fig. 4G). Only CSF/skull interfaces, for example, in frontal regions, are slightly compromised by a more global ringing, presumably from residual fat or high-intensity signal right after slice excitation, and the reversed T2* weighting in k-space for spiral-ins, amplifying high spatial frequencies of the image. The out-part of the spiral-in/out (Fig. 4D,H) constitutes a compromise between spiral-in and high-resolution spiral-out in terms of artifact-level. Its shorter readout of only 20 instead of 60 ms alleviates through-plane dephasing or incomplete B0 correction through inaccurate mapping.
We further quantified the spatial specificity of the high-resolution spiral-out images using contour distance mapping (Fig. 5). This measure is visualized as colored anatomical image contour on top of the mean spiral image (Fig. 5A, depicting subject S3) and shows in general good congruence of functional and structural data (0 to 1.5 voxel contour distance in most voxels). Larger deviations occurred at tissue boundaries with large susceptibility gradients, e.g., close to the frontal sinus (Fig. 5A, left inset, white arrows), as well as areas with pronounced T2* contrast differences between the functional and structural scan, such as subcortical gray matter (Fig. 5A, right inset, yellow arrows).
Overall, in most slices and subjects, the mean distance between corresponding
contours of the structural and functional image varied between 0.5 and 1.5
voxels, with generally better congruence for more superior slices (Fig. 5B).
Prominent outliers (subject 5,6, inferior slices) arose in areas with signal
dropout for the spiral-out image, which was more susceptible to through-plane
dephasing than the structural ME scan due to the difference in echo time (TE
2o ms vs 5-10 ms). For visual cortex in particular (Fig. 6B), the region of
interest for our functional analysis, contour congruencies between 0.5 and 1 mm
were most common, with fewer outliers than in the rest of the brain, driven by
both contrast differences (S2, slice 10) and signal dropouts close to the
sagittal sinus (S2, slice 34). Averaged over the whole volume and all subjects,
the mean gray matter contour distance amounted to
Summarizing the distribution of contour distances over all voxels in the
acquisition volume (Fig. 5D),
Functional sensitivity of the high-resolution spiral-out images is evident at the single-subject level (subject S2) in a differential contrast of both stimulus conditions (+/- ULLR-URLL). The corresponding t-map, overlaid on the mean functional images, reveals expected activation patterns in visual cortex (Fig. 6A). Hemispheric separation of the complementary quarter-field stimulation blocks is visible (left slice), as well as the contrast inversion from inferior to superior slices (leftmost slice vs second from left). Notably, significant activation flips between neighboring voxels occur at the cerebral fissure, suggesting spatial specificity at the voxel level.
This functional specificity is confirmed when overlaying the identical activation maps on the mean ME image as anatomical reference (Fig. 6B), again demonstrating the good alignment of functional and structural data seen in the previous subsection (Figs. 4,5). Clustered activation is almost exclusively constrained to gray matter with no extension into adjacent tissue or skull. Note that no multiple comparison correction was performed for visualization, in order to be more sensitive to such effects, at the expense of occasional false-positive voxels throughout other brain areas.
Gray-matter containment and retinotopic organization of the activation can be further corroborated in the zoomed-in sections of visual cortex for transverse, coronal and sagittal orientation (Fig. 6C). Additionally, we evaluated the ULLR and URLL blocks individually (Fig. 6D), because differential contrasts, by design, do not allow for spatial overlap between significant activation of both conditions. In the individual contrasts, the identified portion of activated visual cortex appears larger, but is still very well restricted to cortical gray matter. Few overlaps exist, and, again, contrast switches between adjacent voxels, pointing to spatial specificity at the prescribed sub-mm resolution.
These findings are reproducible over subjects (Fig. 7). Importantly, similar image quality and geometric congruency are accomplished in all subjects. To verify, we show both the mean spiral and the anatomical ME reference image of the corresponding transverse slice as underlays for the differential activation patterns. Some subjects exhibit more frontal blurring artifacts and dropouts (S5, S6, S7) due to different geometry of the air cavities. Still, the retinotopic organization of visual cortex is recovered in all subjects, as visualized in the zoomed coronal and sagittal views. Existing differences of the specific activation patterns are within the expected range of variability in subject anatomy and task engagement. Quantitatively, peak t-values reach 15.1 on average for the differential contrasts, with a standard deviation of 2.6, i.e. 17 %, over subjects (Table 1). Activation clusters comprise 10371 +/- 2480 voxels (after FWE-multiple comparison correction at the cluster level, p < 0.05), i.e., 6467 +/- 1547 mm3.
Because traditional Gaussian smoothing is frequently omitted in ultra-high resolution fMRI studies (e.g., for laminar fMRI), we assessed its impact on our results. We repeated the statistical analysis for all subjects to create a version of Figure 7 based on unsmoothed data (supplementary material SM 3). For one particular subject (S2), we also juxtaposed spatial characteristics of the statistical t-maps in an animated slide show by varying significance thresholds for smoothed and unsmoothed data, as well as cropping the spiral k-space data to 1 mm resolution before reconstruction (supplementary material SM 4). Overall, spatial smoothing increased overall CNR (higher t-values) and extent of activation clusters that were already discernible in the unsmoothed data. The activation clusters of the smoothed data resemble those of unsmoothed data at lower thresholds, but with fewer single-voxel activation sites. When overlaying activation masks of both differential t-contrasts (+/- ULLR-URLL) after cluster-level multiple comparison correction (p < 0.05 cluster-FWE, cluster-forming threshold: p < 0.001) for this subject directly, we observed two distinct effects of the employed moderate smoothing (FWHM 0.8 mm) (Fig. 8A): on the one hand, cluster extent may increase isotropically by about one voxel (left inset), consistent with a loss in spatial specificity. On the other hand, clusters can expand by several voxels along the cortical ribbon after smoothing (right inset), suggesting that increased sensitivity by averaging of thermal noise can lead to functionally more plausible activation patterns.
To quantify functional specificity, we evaluated the tissue type of all
significantly activated voxels, and assessed the impact of smoothing on this
measure (Fig. 8B), with tissue type based on the unified segmentation results of
the structural data (mean ME). For smoothed data and considering the whole
imaging volume (top left),
We continue to present data from the same subject (S2) as in the high-resolution case, to facilitate comparison. All findings are generalizable over subjects, and we provide mean, SD, SFNR and t-maps of all slices for further dissemination in the supplementary material (SM 2).
Overall, the differential t-contrast maps for the spiral in/out data resemble the activation patterns of the high-resolution spiral-out case (Fig. 9). This holds for all three derived in/out time series, i.e., the separate reconstructions of the in-part and the out-part, as well as their combination in the image domain via signal-weighted averaging (“in/out combined”).
In terms of functional sensitivity, the in/out sequence provides higher peak t-values and cluster extents in the differential t-contrasts compared to the high-resolution spiral-out, as expected due to the larger voxel size and consequential higher SFNR (Table 1). For example, the in-part itself provides a 61 % SFNR increase in gray matter (averaged over subjects), 17 % increased maximum peak t-value, and 56 % increase in significantly activated gray matter volume (Table 1, rightmost column).
Comparing the out- to the in-part of the spirals, SFNR is slightly decreased in the out-part (8 %), while the situation is reversed for the t-maps, with 2 % increase in peak t-value and 14 % increase in cluster extent, compared to the spiral-in. This suggests that higher T2*-sensitivity of the spiral-out causes both effects, by both amplifying signal dropouts and BOLD signal.
The signal-weighted echo combination (eq. (6), (Glover and Thomason, 2004)) provides the highest functional sensitivity of the three in/out time-series, having a 25 % increased SFNR compared to the in-part, and 37 % increase compared to the out-part. This translates into an average increase in peak t-value of 2 % and significant cluster extent of 21 %, compared to the out-part alone. This is in line with previous findings for high-resolution multi-shot spiral data (Singh et al., 2018) at 3 T, which also reported contrast-to-noise ratio (CNR) increases for signal-weighted spiral in/out combinations on the order of 25 %. However, it falls somewhat short of the 54 % increase in CNR reported originally for low-resolution single-shot spiral in/out combination (Glover and Thomason, 2004, p. 866).
In terms of spatial specificity, all activation patterns exhibit a good congruency to the anatomical reference, as evident from a close-up overlaid onto the mean ME image (Fig. 9D). In general, this visualization confirms the overall impression that the echo combination increases CNR throughout visual cortex, rather than just in regions of higher dephasing. Remarkably, there seem to be more false positive clusters for the spiral-in than in all other spiral variants (Fig. 9A), in particular close to the temporal muscle, presumably due to the ringing mentioned above.
In this work, we demonstrated that recent advances in high-resolution, single-shot spiral imaging (Engel et al., 2018) can be deployed to fMRI. The typical drawbacks of spiral fMRI, which have so far limited its routine use, were overcome by an expanded signal model, accurate measurement of its components, and corresponding iterative image reconstruction (Barmet et al., 2005; Pruessmann et al., 2001; Wilm et al., 2011).
Specifically, time series of high image quality and stability were obtained that exhibited geometric congruency to anatomical scans without the need for post-hoc distortion correction. Notably, also the corresponding phase images exhibit high raw data quality (without any postprocessing, e.g., phase unwrapping), and suggest the suitability of spiral acquisition for novel phase- or complex-value based fMRI analysis workflows (Balla et al., 2014; Bianciardi et al., 2014; Calhoun et al., 2002; Menon, 2002).
The functional sensitivity of spiral readouts was confirmed by observing typical activation patterns in response to an established visual quarter-field stimulation. While a consensus on how to assess spatial specificity for fMRI is lacking, several indicators point to a localization capability in the sub-mm range for our data. First, the distance of gray matter contours in spiral and structural MRI data were at most one voxel (0.8 mm) apart in the vast majority of voxels per subject (76 %). Second, the activation patterns of different stimulus conditions could be discriminated in neighboring voxels of 0.8 mm nominal resolution (Fig. 6). Third, the vast majority (75 %) of significant activation sites were contained within gray matter or at its boundaries, suggesting a limited impact of artifactual blurring.
Finally, we demonstrated the versatility of this approach to spiral fMRI with a combined in/out readout at a more typical resolution (1.5 mm). Here, the high acquisition efficiency of the spiral allowed to measure two images per shot, increasing CNR by about 20 %. The observed discrepancy to previously reported gains of more than 50 % for signal-weighted echo combination (Glover and Thomason, 2004) is in line with recent spiral fMRI studies (Singh et al., 2018). It might result from higher target resolution and static off-resonance correction employed in our study and (Singh et al., 2018), compared to the original work. The increased image congruency and smaller dephasing effects between the in- and out-part compared to low-resolution spirals may reduce the impact of echo combination. Still, more sophisticated combination of echo images (Glover and Thomason, 2004), or of k-space data during reconstruction (Jung et al., 2013) could result in further SNR increases.
In summary, the presented advances render spiral fMRI an attractive sampling scheme that delivers on the long-time postulate of high acquisition efficiency without compromising image quality. Here, the spatiotemporal application domain of fMRI on a standard gradient system was enhanced by acquiring a 230x230x36 mm FOV brain image at 0.8 mm nominal in-plane resolution (i.e., a matrix size of 288x288x36) while maintaining a TR typical for high-resolution fMRI (3.3 s). This corresponds to an acquisition efficiency of about 900,000 resolved voxels per second.
To our knowledge, this is the highest acquisition efficiency reported for 2D spiral fMRI to date (see Table SM 5 for a comparison of sequence parameters in several spiral fMRI studies), as well as the first high-resolution spiral fMRI study at ultra-high field. In combination with the presented evidence for geometric accuracy, this makes the presented spiral-out sequence an attractive candidate for high-resolution applications of fMRI, studying the functional sub-organization of cortex, e.g., in laminae or columns (Cheng et al., 2001; Feinberg et al., 2018; Fracasso et al., 2016; Huber et al., 2017a; Kok et al., 2016; Koopmans et al., 2010; Lawrence et al., 2018; Martino et al., 2015; Muckli et al., 2015; Siero et al., 2011; Uğurbil et al., 2013; Yacoub et al., 2008).
The claim of high acquisition efficiency for fMRI hinges on whether the acquired voxels effectively resolve distinct activation. This question of effective functional resolution arises for any fMRI protocol and comprises global aspects, such as PSF broadening, as well as more localized effects concerning geometric congruency and spatial specificity of the activation mapping.
The circular k-space coverage of spirals leads to a broadening of the PSF main lobe by 17 % compared to Cartesian k-space coverage in EPI (1.4 vs 1.2 times the voxel size (Qin, 2012)). Furthermore, any sequence with long readout duration encounters considerable T2* signal decay along the trajectory, which manifests as a filter in image domain. For spiral-in images, this emphasizes higher spatial frequencies, while the effect on spiral-out images is reversed, leading to blurring. Based on typical brain tissue relaxation times at 7T, we adapted previous simulations of this effect for a similar high-resolution spiral-out protocol (Fig. 7 in Engel et al., 2018). There are diminishing returns for investing more acquisition time to achieve higher in-plane resolution, but a net gain remains at our chosen readout duration of 60 ms. Effectively, the FWHM of the PSF due to T2* blurring corresponds to a voxel size smaller than 1 mm for the targeted 0.8 mm nominal resolution, while at 40 ms readout duration actual voxels are larger than 1.1 mm for a targeted 1 mm nominal resolution. Still, choosing shorter readouts with slightly coarser resolution in favor of sampling more slices within the given TR might deliver overall higher acquisition efficiency in this case. Finally, static B0 inhomogeneity also manifests as spatially varying blurring or ringing in spiral imaging, because off-resonance induces broadening of the PSF main lobe, as well as amplification of its side lobes (Bernstein et al., 2004, Chap. 17; Fig. 6 in Engel et al., 2018; Man et al., 1997). As long as B0 inhomogeneity is properly mapped and included in the signal model, this effect is mitigated by the iterative image reconstruction utilized in this work.
In our experimental data, we found that the spatial specificity of spiral fMRI is very high in about 75-80 % of the voxels, as indicated by both contour distance mapping (Fig. 5) and gray matter containment of activation (Fig. 8). While these quantifications provide rough estimates of functional spatial specificity in fMRI, there is no consensus on such quantification in the community, and the absolute values reported here are hard to compare to previous work. We hope that through our sharing of the code, this methodology may provide future reference points. Furthermore, the utilized measures are in themselves imperfect assessments of spatial specificity, and might underestimate the achieved spatial specificity in our data. First of all, both analyses relied on unified segmentation (Ashburner and Friston, 2005) to extract tissue probability maps, which in principle should work contrast-independently. However, the employed default parameter settings may not be optimal to model bias fields at 7 T or voxel intensity distributions of multi-echo GRE scans with only partial brain coverage. Additionally, because the contrast in our functional and structural scans were not equivalent (e.g., TE 20 vs 10 ms), contour distance mapping reflects, to a certain extent, differences in contrast rather than geometry.
Finally, relying on a perfect retinotopic organization of visual cortex for assessing functional spatial specificity has also limitations. For example, receptive fields can cross the vertical meridian (especially in higher visual areas), such that differential contrasts of quarterfield stimulation may not flip between adjacent voxels (Fig. 6C). Similarly, overlapping voxel activation or single-voxel “false positives” (Fig. 6D) may indicate imperfections in the visual field leading to non-compact retinotopic representations, rather than losses in spatial specificity.
Another contentious point concerning spatial specificity is our choice of moderately smoothing the data (FWHM 0.8 mm) before statistical parametric mapping. While smoothing is a standard preprocessing step in the majority of fMRI studies, high-resolution applications, such as retinotopic mapping or layered fMRI analyses, frequently abstain from it or use more spatially informed averaging methods (Blazejewska et al., 2019). From a conceptual point of view, smoothing alters the target PSF of the imaging process in multiple ways that affect the effective resolution: On the one hand, a Gaussian filter, as employed here, broadens the PSF main lobe (PSF) reducing spatial specificity. On the other hand, it suppresses PSF side lobes and thus contamination by remote locations, which enhances spatial specificity. The opposite is true for an edge or high-pass filter, as implemented, for example, by multiplying the inverse of the T2* decay curve onto the k-space data. Overall, resolution can be re-negotiated by appropriate filtering, which has to be adapted to the specific application, in order to provide optimal specificity.
For our spiral data in particular, the decision to smooth was governed by another goal of filtering, namely, recovering sensitivity. Theoretically, sensitivity is maximized by a matched filter resembling the spatial activation extent, which is traditionally assumed to be a Gaussian for fMRI (Friston, 2007, Chap. 2; Kasper et al., 2014), but has, to our knowledge, not been determined for the ultra-high-resolution regime in question here. Our choice of smoothing with an FWHM equivalent to the voxel size (instead of 2-3 times the voxel size as in standard fMRI) therefore constitutes a compromise between sensitivity and specificity, motivated by the noise levels in our raw data, the short run duration (5.5 min) and by having only one run acquired per subject and spiral sequence. This hampered our ability to quantify whether the investment into longer readouts of nominal 0.8 mm resolution, compared to 1 mm, translated into more spatially specific activation (supplementary material SM 4). More temporal averaging via longer and more numerous functional runs would allow to address this important research question in future studies.
Increasing acquisition efficiency for high-resolution single-shot spirals while maintaining depiction quality, as presented here, resulted from the favorable interplay of the expanded signal model components: the encoding field dynamics with long-readout spirals, static B0 inhomogeneity characterization, and parallel imaging with iterative reconstruction enabling undersampling.
For deploying this advanced spiral functional imaging to other sites and systems, it is important to evaluate how individual aspects of the approach contribute to its overall performance, and to assess the generalizability of our findings. This includes both the impact of model and system components, as well as the utilized methodology for their characterization, in relation to possible alternatives and extensions.
In terms of availability, the concurrent field monitoring hardware employed in our approach (Barmet et al., 2008; Engel et al., 2018; Kasper et al., 2018; Wilm et al., 2017, 2011), is probably the scarcest resource across sites. It serves to characterize both the reproducible and irreproducible imperfections of the encoding magnetic fields.
For reproducible field effects, such as the actual spiral trajectory performed by the system and its induced eddy currents, previous work has shown that their characterization is often required to avoid severe image artifacts (Engel et al., 2018, Fig. 6; Vannesjo et al., 2016). This, however, might vary between systems, as successful spiral image reconstructions based on nominal trajectories have been reported (Kurban et al., 2019; Singh et al., 2018). If image artifacts arise from reproducible trajectory imperfections, they could be measured without concurrent field monitoring hardware by calibration approaches in a separate scan session (Bhavsar et al., 2014; Duyn et al., 1998; Robison et al., 2019, 2010). For more flexibility, the gradient impulse response function (GIRF) to arbitrary input trajectories can be modelled from such data under linear-time invariant system assumptions (Addy et al., 2012; Campbell-Washburn et al., 2016; Rahmer et al., 2019; Vannesjo et al., 2014, 2013). The required field measurements for these calibrations may either rely on dedicated NMR-probe based field cameras (Barmet et al., 2008; Vannesjo et al., 2014, 2013; Zanche et al., 2008) or on off-the-shelf NMR phantoms (Addy et al., 2012; Duyn et al., 1998; Rahmer et al., 2019), with certain trade-offs to measurement precision and acquisition duration (Graedel et al., 2017).
For the dynamic field effects, induced by the system (e.g., drifts through gradient heating), as well as the subject (e.g., fluctuations with the breathing cycle or limb motion), few studies have analyzed their impact on spiral fMRI time series (Pfeuffer et al., 2002). In principle, the concurrent field monitoring and reconstruction employed here incorporated changes of global off-resonance and k-space with the bandwidth of the trajectory measurement of about 4 Hz (monitoring every third slice). As we did not observe any conspicuous problems in the time series statistics, for example, SFNR drops, nor any indication of time-dependent blurring, which would be the spiral equivalent to apparent motion in phase encoding direction observed in EPI (Bollmann et al., 2017; Power et al., 2019), this approach presumably addressed the majority of field fluctuations present in our data. This is in line with previous results of breathing-induced field fluctuations reported at 7T in spiral fMRI of only a few Hz for healthy subjects and normal breathing (Pfeuffer et al., 2002). An in-depth analysis of these effects is beyond the scope of this paper, as it would, for example, entail quantitative comparisons with nominal or GIRF reconstructions (Vannesjo et al., 2016), as well as simulating the impact of different measured field components on image time series, similar to work previously conducted for EPI (Bollmann et al., 2017; Kasper et al., 2015). As we do believe that this investigation is relevant to the neuroimaging community, we provide the field dynamics of all spiral-out fMRI runs in ISMRMRD format for further scrutiny.
Note, however, that the dataset in itself might not be representative for
assessing the utility of concurrent field monitoring for spiral fMRI. In terms
of system fluctuations, we did have a challenging gradient duty-cycle (with
gradients switched on during 70 % of the sequence, 50 % of the time at amplitude
maximum), leading to substantial heating of 15 degrees throughout the 5.5 min
high-resolution spiral-out sequence, as measured using 5 optical sensors cast
into the gradient coil (Dietrich et al., 2016b). This actually limited the
duration of our functional runs. While system fluctuations might thus be
particularly pronounced in our data, the subject-induced fluctuations will be
moderate, because all volunteers were young, healthy individuals instructed to
lie still throughout the session. This is reflected in the small mean framewise
displacement (FD) encountered in all subjects (mean FD and standard deviation
over subjects
If dynamic field effects constitute a significant artifact and noise source for spiral fMRI time series, in lieu of field monitoring, alternative correction methods comprise dynamic off-resonance updates or higher-order field navigators (Pfeuffer et al., 2002; Splitthoff and Zaitsev, 2009), as well as gradient response models that incorporate time-courses of gradient coil temperature (Dietrich et al., 2016b; Stich et al., 2020) or current measurements (Nussbaum et al., 2019; Rahmer et al., 2021).
To characterize static B0 inhomogeneity, we acquired a Cartesian multi-echo gradient echo scan with rather high resolution (1 mm). Including this information in the signal model has previously been crucial to maintain spatial specificity in spiral imaging at 7T (Engel et al., 2018, Fig. 6; Kasper et al., 2018, Fig. 7).
The B0 maps obtained in this work exhibited considerable inhomogeneity, even after 3rd order shimming of the targeted 4 cm oblique-transverse slab of the brain including the visual cortex. For the B0 map of a single subject (SPIFI_0007) provided in the accompanying Data In Brief article (see "Code and data availability" section), 20 % of brain voxels were more than 50 Hz off-resonant, which – if uncorrected for – would incur blurring with FWHMs of several voxels.
Thus, some form of static B0 inhomogeneity correction seems indispensable for providing high spiral image quality, and to correct this at the reconstruction stage via inclusion into the expanded signal model proved sufficient for most of the imaged brain slices. However, localized blurring, distortion and ringing artifacts remained at cortex boundaries close to the skull or air cavities, most prominently in orbitofrontal regions, and in the temporal lobe, above the ear canals, as well as in more inferior slices, particularly in the brainstem. Consequently, such regions might exhibit less sensitive and spatially less defined activation patterns than the ones in visual cortex evaluated here.
We did not evaluate whether our particular choice of B0 map resolution or processing contributed to the accomplished image quality or its limitations. Alternative methods to determine B0 may provide similar results at reduced scan time, for example, slightly varying TEs during a spiral image time series to estimate the B0 map from their phase differences directly (Glover and Law, 2001; Singh et al., 2018) or joint estimation of B0 and image from the spiral data itself (Fessler, 2010; Hernando et al., 2008; Patzig et al., 2020). These methods also allow to regularly update B0 maps during long fMRI runs, increasing the alignment to the spiral acquisition geometry in case of subject motion.
To enable single-shot imaging for maximum acquisition efficiency, we also relied on the coil sensitivity profiles for spatial encoding, i.e., parallel imaging. Spiral trajectories are particularly suited for this form of acceleration, because they possess favorable behavior in terms of spatial noise amplification by the coil geometry factor, allowing for higher k-space undersampling (Larkman, 2007; Lee et al., 2021).
For our data with in-plane acceleration factors of 4 using a 32-channel receive array at 7 T this was confirmed through the SD maps of the time series data, which did not exhibit spatially structured residual aliasing or noise patterns. This also points to the robustness of the reconstruction to motion-induced mismatch of measured and actual coil sensitivities.
Compared to Nyquist-sampled spiral data, which could be reconstructed via gridding and conjugate phase correction for static off-resonance effects (Singh et al., 2018 and references therein), parallel imaging of non-Cartesian trajectories typically necessitates iterative reconstruction schemes (Heidemann et al., 2006; Lustig and Pauly, 2010; Pruessmann et al., 2001; Weiger et al., 2002; Wright et al., 2014). Depending on the number of iterations and precision of the off-resonance correction, these algorithms are one or two orders of magnitude slower than direct reconstruction methods. Note, however, that the reconstruction times reported here will not present a general hurdle for deployment, because our Matlab code was not optimized for speed. The numerous matrix-vector multiplications (eq. 5) burden the CG algorithm most, and an implementation on graphical processing units (GPUs) will significantly accelerate reconstruction. High-performance implementations of the conjugate gradient iterative reconstruction algorithm, including off-resonance correction, are publicly available in different MR reconstruction packages, and we have successfully tested reconstruction of the example data presented here in MRIReco.jl (Knopp and Grosser, 2021), written in the modern scientific programming language Julia (Bezanson et al., 2017).
Finally, the availability of an ultra-high field system may be seen as a limitation for the advances presented here. We implemented the spiral sequences at 7T, which has shown particular utility for high-resolution functional MRI due to its superlinear increase in BOLD CNR (Uludağ and Blinder, 2018). From an image reconstruction perspective, this is a challenging scenario, because both static and dynamic field perturbations are exacerbated at ultra-high field and deteriorate conditioning of the expanded signal model. Thus, the adoption of the presented advances in spiral fMRI to lower field strengths (e.g., 3T) not only seems straightforward and worthwhile, but also offers benefits. For example, spiral readouts could be prolonged in light of the more benign field perturbations, mitigating the lower CNR while maintaining high image quality.
Our gradient system, on the other hand, had standard specifications, available on most sites (utilized gradient amplitude 31 mT/m, slew rate 160 T/m/s). Already here, spiral trajectories offered reduced readout times of 19 % compared to EPI due to their higher average speed covering k-space. Because the last 80 % (45 ms) of our high-resolution spiral gradient waveform were amplitude-limited (Fig. 1, black waveform), this acceleration could be considerably increased by dedicated gradient hardware with higher maximum gradient strength, e.g., high-performance whole-body “connectome” gradient coils (Kimmlingen et al., 2012) or insert gradients for head imaging (Foo et al., 2018; Weiger et al., 2018).
This work focused on two-dimensional, slice-selective spiral BOLD imaging. Simultaneous multi-slice (SMS) or 3D excitation schemes offer a complementary means of acceleration, by extending sensitivity encoding to the third encoding (slice) dimension, as, e.g., in stack-of-spiral trajectories (Deng et al., 2016; Engel et al., 2021; Zahneisen et al., 2014), which also provides SNR benefits (Poser et al., 2010). The expanded signal model and image reconstruction framework employed here, apart from the 2D-specific simplifications, are equally applicable to this scenario (Engel et al., 2021; Pruessmann et al., 2001; Zahneisen et al., 2015).
Furthermore, the successful deployment of the in/out spirals here suggests the feasibility of other dual-echo variants, such as out-out or in-in acquisition schemes. In particular, recent correction methods for physiologically or motion-induced noise that rest on multi-echo acquisition (Kundu et al., 2012; Power et al., 2018) could profit considerably from spiral-out readouts: compared to EPI, the shorter minimum TE provides first-echo images with reduced T2*- weighting and should enhance disentanglement of BOLD- and non-BOLD related signal fluctuations.
Beyond BOLD, the adaptation of single-shot spiral acquisition for other time series readouts seems promising. In particular fMRI modalities with different contrast preparation (Huber et al., 2017b), such as blood-flow sensitive ASL (Detre et al., 2012, 1992), and blood-volume sensitive VASO (Huber et al., 2018; Lu et al., 2013, 2003) benefit from the shorter TEs offered by spiral-out readouts. These sequences do not rely on T2* decay for functional sensitivity, and thus minimizing TE leads to considerable CNR gains (Cavusoglu et al., 2017; Chang et al., 2017).
This work was supported by the NCCR “Neural Plasticity and Repair” at ETH Zurich and the University of Zurich (KPP, KES), the Clinical Research Priority Program of the University of Zurich, CRPP “Pain” (MMS, KES), the René and Susanne Braginsky Foundation (KES), the University of Zurich (KES) and the Oxford-Brain@McGill-ZNZ Partnership in the Neurosciences (NNG, OMZPN/2015/1/3). Technical support from Philips Healthcare, Best, The Netherlands, is gratefully acknowledged. The Wellcome Centre for Integrative Neuroimaging is supported by core funding from the Wellcome Trust (203139/Z/16/Z).
We thank the reviewers for very constructive feedback that helped to considerably improve our manuscript, in particular the concrete suggestions regarding quantification of spatial accuracy of our spiral data via contour distance mapping. We are grateful to Roger Luechinger for technical support with scanning and data curation.
At the time of submission, Christoph Barmet and Bertram J. Wilm are employees of Skope Magnetic Resonance Technologies. Klaas P. Pruessmann holds a research agreement with and receives research support from Philips. He is a shareholder of Gyrotools LLC.
Utilized 2D single-shot spiral acquisitions (R = 4 undersampling): High-resolution single-shot spiral-out (nominal resolution 0.8 mm, black) and spiral in/out trajectory (1.5 mm resolution, blue). Depicted are the gradient waveforms (Gx,Gy,Gz) as well as RF excitation (TX) and ADC sampling intervals (AQ) for both the 1H head coil and the 19F field probes used to monitor the trajectories and other concurrent encoding fields. Field probe excitation and acquisition start a few milliseconds before the spiral readout gradient waveforms.
Overview of image quality for high-resolution (0.8 mm) single-shot spiral-out acquisition. (A) 8 oblique-transverse slices (of 36) depicting the time-series magnitude mean of one functional run (subject S7, 100 volumes). (B) Single-volume magnitude images for slices corresponding to lower row of (A). (C) Mean phase image over one run, without any post-processing, for slices corresponding to lower row of (A).
Characterization of image time series fluctuations over 1 spiral-out run (95 volumes, discarding first five). (A) Signal-to-Noise Fluctuation Ratio (SFNR) image for same slices as in Fig 2. Rather homogeneous, exhibiting sufficient SFNR levels. (B) Standard deviation (SD) image over time. Regions of high fluctuation mainly include pulsatile areas close to ventricles or major blood vessels, and cortex/CSF interfaces. (C) Coefficient of Variation (CoV) image. Inverse of (A), highlighting regions of high fluctuations relative to their respective mean. Vascularized/CSF regions appear prominently, as well as the internal capsule, due to its reduced average signal level.
Image quality and geometric accuracy of spiral images, reconstructed with the expanded signal model. (A) Anatomical Reference: Mean multi-echo (ME) spin-warp image (1 mm resolution) (B) High-resolution (0.8 mm) spiral-out; (C) In-part of spiral in/out (1.5 mm); (D) Out-part of spiral in/out (1.5 mm). (E) B0-map computed from (A). (F-H) Overlay of isoline contour edges from (A) onto (B)-(D).
Depicted are the mean images of a single run (subject S2, top row, B-D). The mean ME image (A), used to compute SENSE- and B0-map (E) for the expanded signal model, provides the anatomical reference via its contours (red lines), overlaid onto the different spiral variants (bottom row, F-H). Arrows indicate residual geometric incongruence by through-plane dephasing (white) or incomplete B0 mapping and correction (yellow) in the spiral-out, which are reduced in the out-part and absent in the in-part of the spiral-in/out sequence.
Quantification of spatial specificity in spiral images via contour distance
mapping.
(A) Gray matter contours extracted from tissue probability maps (threshold 90 %)
of the T2*-weighted structural image (TE 10 ms, echo 6 of ME scan), overlaid
onto mean high-resolution spiral-out image (subject S3). Color coding reflects
distance to corresponding contour in segmented spiral image. Contour
discrepancies are prevalent at tissue interfaces with high susceptibility
gradients (left inset, white arrows), as well as areas with pronounced T2*
contrast differences (right inset, yellow arrows). (B) Mean contour distance per
slice for different subjects (averaged over all contours within each slice for
the whole imaging volume). Average over subjects and slices:
Visual Activation Maps of high-resolution (0.8 mm) spiral-out fMRI for a single subject (S2). Representative stimuli of both conditions (ULLR and URLL) are displayed at the top. (A) Overlay of differential t-contrast maps (p < 0.001 uncorrected) on transverse slices of mean spiral image (hot colormap: URLL-ULLR, cool colormap: ULLR-URLL). (B) Same contrast maps as in (A), overlaid on mean ME image as anatomical reference. (C) Zoomed-in sections of differential t-contrast maps in different orientations: transverse (top), coronal (middle) and sagittal (bottom, left (L) and right (R) hemisphere). (D) t-contrast maps for individual conditions (blue: ULLR, yellow: URLL), showing more widespread activation and high spatial specificity, i.e., little spatial overlap (green).
Mean spiral images and activation maps over subjects (S2-S7) for high-resolution spiral-out fMRI. For each subject, the following 4 sections are displayed, with the mean ME image as anatomical underlay: transverse, coronal and sagittal slice (for left (L) and right (R) hemisphere), each chosen for the maximum number of activated voxels (over both differential statistical t-contrasts, p < 0.001 uncorrected). To assess raw spiral data quality, the corresponding mean functional image is displayed side-by-side to the anatomical transverse slice as an alternative underlay.
Spatial specificity of functional activation for high-resolution spiral-out fMRI. Analyses are based on significant voxels over both differential t-contrasts (+/- ULLR-URLL, p < 0.05 cluster-FWE corrected, cluster-forming threshold: p < 0.001). (A) Comparison of activation extent in smoothed (FWHM 0.8 mm) and unsmoothed data in a single subject (S2). Masks of all significant voxels are overlaid for both t-contrasts based on smoothed data (blue/red mask) as well as unsmoothed data (cyan/yellow masks). (B) Percentage of significant voxels located in relevant tissue types, analyzed for smoothed (top row) and unsmoothed (bottom row) data, as well as within whole imaging volume (left) and restricted to a mask of early visual cortex (right) . Tissue types were identified by unified segmentation of the structural (mean multi-echo) image, with 60 % exceedance threshold for individual tissue classes (GM: gray matter, WM: white matter, CSF: cerebrospinal fluid) and 30 % each for interfaces (Pial surface (GM/CSF), WM/GM surface), with the remaining voxels categorized as ambiguous. On average, irrespective of smoothing and ROI, the majority (74-78 %) of activation was contained in gray matter or adjacent surfaces, with 6 % and 13-17 % residing in majority white matter and CSF voxels, respectively. Gray matter containment was highest when smoothing the data and restricting the analysis to the visual cortex, and lowest in the unsmoothed data considered within the whole imaging volume.
Visual Activation Maps of spiral in/out (1.5mm) fMRI run for a single subject (S2, as in Fig. 6). (A-C) Displayed are the differential t-contrast maps (p < 0.001 uncorrected) on transverse slices of the respective mean spiral images (hot colormap: URLL-ULLR, cool colormap: ULLR-URLL), based on: (A) Spiral Images reconstructed from in-part of the trajectory. (B) Spiral images reconstructed from the out-part of the trajectory. (C) Signal-weighted combination (eq. (6), (Glover and Thomason, 2004)) of images in (A) and (B). (D) Zoomed view of activation maps in leftmost slice of (A)-(C), overlaid on anatomical reference image (mean ME).
Quantification of temporal stability and functional sensitivity of all spiral fMRI sequences. For the signal-to-fluctuation-noise ratio (SFNR, eq. (5)), the table contains mean +/- SD in a gray matter ROI over the whole imaging volume. For the t-contrast SPMs, peak t-value and number of significant voxels over both differential contrasts (+/- ULLR-URLL) are reported (p < 0.05 FWE-corrected for multiple comparisons at the cluster level with a cluster-forming threshold of p < 0.001). The last column shows relative increases to the previous sequence, i.e., the one reported in the sub-table directly above. Since resolutions differ between spiral-out (0.8 mm) and spiral in/out (1.5 mm), we compare activated volume instead of voxel count.
Subjects | Gain vs | |||||||||
---|---|---|---|---|---|---|---|---|---|---|
Measure | S2 | S3 | S4 | S5 | S6 | S7 | Mean | SD | previous spiral | high-res spiral out |
high-resolution spiral out | ||||||||||
SFNR_mean | 14.1 | 17.2 | 14.9 | 15.7 | 14.7 | 15.2 | 15.3 | 1.1 | - | - |
SFNR_SD | 5.0 | 5.4 | 4.6 | 5.0 | 4.7 | 4.9 | 4.9 | 0.3 | ||
SPM_T_max | 12.9 | 15.2 | 16.2 | 17.5 | 11.1 | 17.6 | 15.1 | 2.6 | - | - |
SPM_T_nVoxels | 10957 | 6157 | 11673 | 12202 | 8682 | 12553 | 10371 | 2480 | ||
SPM_T_volume (mm3) | 6832 | 3839 | 7279 | 7609 | 5414 | 7828 | 6467 | 1547 | - | - |
in-part spiral in/out | ||||||||||
SFNR_mean | 22.7 | 28.6 | 23.5 | 26.6 | 22.9 | 23.9 | 24.7 | 2.4 | 61% | 61% |
SFNR_SD | 7.3 | 8.2 | 7.6 | 7.8 | 7.3 | 8.2 | 7.7 | 0.4 | ||
SPM_T_max | 16.6 | 19.5 | 23.5 | 15.1 | 15.3 | 15.9 | 17.7 | 3.3 | 17% | 17% |
SPM_T_nVoxels | 10097 | 4863 | 10221 | 6470 | 8157 | 2816 | 7104 | 2953 | ||
SPM_T_volume (mm3) | 14363 | 6918 | 14540 | 9204 | 11604 | 4006 | 10106 | 4200 | 56% | 56% |
out-part spiral in/out | ||||||||||
SFNR_mean | 20.4 | 26.4 | 21.4 | 24.3 | 21.0 | 22.9 | 22.7 | 2.3 | -8% | 48% |
SFNR_SD | 7.3 | 8.4 | 7.4 | 8.0 | 7.7 | 7.9 | 7.8 | 0.4 | ||
SPM_T_max | 17.8 | 20.2 | 18.2 | 20.1 | 17.2 | 14.4 | 18.0 | 2.1 | 2% | 19% |
SPM_T_nVoxels | 8976 | 6087 | 10661 | 7923 | 11606 | 3312 | 8094 | 3054 | ||
SPM_T_volume (mm3) | 12769 | 8659 | 15166 | 11271 | 16510 | 4711 | 11514 | 4344 | 14% | 78% |
combined spiral in/out | ||||||||||
SFNR_mean | 28.3 | 35.9 | 29.5 | 33.6 | 28.7 | 30.3 | 31.1 | 3.0 | 37% | 103% |
SFNR_SD | 9.4 | 10.9 | 10.0 | 10.4 | 9.9 | 10.7 | 10.2 | 0.5 | ||
SPM_T_max | 18.4 | 19.7 | 20.3 | 17.8 | 15.8 | 18.3 | 18.4 | 1.6 | 2% | 22% |
SPM_T_nVoxels | 11971 | 6575 | 13254 | 9331 | 11665 | 6082 | 9813 | 2985 | ||
SPM_T_volume (mm3) | 17029 | 9353 | 18854 | 13274 | 16594 | 8652 | 13959 | 4247 | 21% | 116% |
Figure collection of high-resolution (0.8 mm) spiral fMRI data. For every subject, 8 figures are shown, each depicting all 36 slices of the acquisition (top left = most inferior slice, bottom right = most superior slice). All underlay images are based on the functional time series after realignment, and the overlaid t-map was computed from the preprocessed data which included smoothing as well (FWHM 0.8 mm). Order of the figures (1) First volume of spiral fMRI time series; (2) Time-series magnitude mean of spiral fMRI run; (3) Bias-field corrected version of (2); (4) Signal-to-Noise Fluctuation Ratio (SFNR) image of spiral fMRI run (display range 0-30); (5) Standard deviation (SD) image over time of the same run (display range 0-0.05); (6) Coefficient of Variation (CoV) image of the same run, inverse of (4) (display range 0-0.2); (7) Magnitude-mean spiral fMRI image, overlaid with edges of anatomical reference (mean multi-echo spin warp); (8) Overlay of differential t-contrast maps (p < 0.001 uncorrected) on transverse slices of mean spiral image (hot colormap: URLL-ULLR, cool colormap: ULLR-URLL, display range t=3.2-8).
Figure collection of spiral in/out fMRI data (1.5 mm resolution). For every subject, 24=3x8 figures are shown, i.e., 8 per set of spiral-in, spiral-out, and combined in/out images. Each figure depicts all 36 slices of the acquisition (top left = most inferior slice, bottom right = most superior slice). All underlay images are based on the functional time series after realignment, and the overlaid t-map was computed from the preprocessed data which included smoothing as well (FWHM 0.8 mm). Order of the figures: (1) First volume of spiral fMRI time series; (2) Time-series magnitude mean of spiral fMRI run; (3) Bias-field corrected version of (2); (4) Signal-to-Noise Fluctuation Ratio (SFNR) image of spiral fMRI run (display range 0-30); (5) Standard deviation (SD) image over time of the same run (display range 0-0.05); (6) Coefficient of Variation (CoV) image of the same run, inverse of (4) (display range 0-0.2); (7) Magnitude-mean spiral fMRI image, overlaid with edges of anatomical reference (mean multi-echo spin warp); (8) Overlay of differential t-contrast maps (p < 0.001 uncorrected) on transverse slices of mean spiral image (hot colormap: URLL-ULLR, cool colormap: ULLR-URLL, display range t=3.2-8).
Mean spiral images and activation maps over subjects (S2-S7) for high-resolution spiral-out fMRI, based on unsmoothed data (See Fig. 7 for same visualization using smoothed data). For each subject, the following 4 sections are displayed, with the mean ME image as anatomical underlay: transverse, coronal and sagittal slice (for left (L) and right (R) hemisphere), each chosen for the maximum number of activated voxels (over both differential statistical t-contrasts, p < 0.001 uncorrected). To assess raw spiral data quality, the corresponding mean functional image is displayed side-by-side to the anatomical transverse slice as an alternative underlay.
Impact of readout duration and spatial smoothing on activation maps in high-resolution spiral-out fMRI data. The animated slideshow toggles between t-maps of the differential contrast (+/- ULLR-URLL), overlaid on the structural image (mean ME) in a single subject (S2): (1) T-map based on smoothed data (FWHM 0.8 mm), thresholded at t > 3.2, (p < 0.001 uncorrected), as presented throughout the other figures, (2) T-map based on unsmoothed data, same thresholding (t > 3.2), and (3) T-map based on unsmoothed data, with more liberal thresholding (t > 2.4, p < 0.01 uncorrected). (4-6) as (1-3), but with spiral-out fMRI k-space data retrospectively cropped to 1 mm resolution before image reconstruction.
(1-3) illustrate that moderate spatial smoothing elevates overall CNR (higher t-values) and increases the spatial extent of existing activation clusters in the unsmoothed data. The activation clusters of the smoothed data resemble those in the unsmoothed data at lower thresholds, but without the single-voxel false-positives. This suggests that the averaging of thermal noise via smoothing delivers an implicit cluster size correction.
(4-6) illustrate that cropping the spiral readout to 1 mm resolution yields similar activation maps to using the full 0.8 mm readout, both before and after smoothing. This indicates that the assessment of spatial specificity is SNR-limited for the short functional runs (5.5 min) investigated here.
Comparison of acquisition efficiency in published 2D spiral fMRI studies. Nominal acquisition efficiency, computed as resolved voxels per unit time (i.e., matrix size per volume TR) is compared for several publications referenced in this manuscript. The combination of single-shot sequences with long readouts and parallel imaging, as utilized in this study, achieves the highest acquisition efficiency.
resolution [mm] | FOV [mm] | matrix size | inter-leaves | TR / interleaf | volume TR | acquisition efficiency | ||||
---|---|---|---|---|---|---|---|---|---|---|
Publication | xy | z | xy | z | xy | z | [s] | [s] | [1000 voxels/s] | |
(Glover and Lai, 1998) | 2.4 | 5.0 | 220 | 30 | 90 | 6 | 1 | 1 | 1.0 | 49 |
(Weiger et al., 2002) | 3 | 5.0 | 240 | 120 | 80 | 24 | 1 | 2.5 | 2.5 | 61 |
(Ress et al., 2007) | 0.7 | 0.9 | 100 | 10 | 141 | 11 | 2 | 1 | 2 | 109 |
0.6 | 0.9 | 100 | 7 | 160 | 8 | 3 | 0.7 | 2 | 102 | |
(Preston et al., 2009; Wolosin et al., 2012; Zeithamova et al., 2016) | 1.7 | 3.0 | 220 | 66 | 129 | 22 | 4 | 1.0 | 4 | 92 |
(Katyal et al., 2010) | 1.2 | 1.2 | 170 | 10 | 142 | 8 | 3 | 1.0 | 3 | 54 |
(Chang and Glover, 2011) | 1.7 | 4.0 | 220 | 96 | 128 | 24 | 1 | 2.5 | 2.5 | 157 |
(Jung et al., 2013) | 1 | 2.0 | 256 | 30 | 256 | 15 | 4 | 1.0 | 4 | 246 |
1.5 | 1.5 | 240 | 36 | 160 | 24 | 2 | 1.5 | 3 | 205 | |
(Kim and Ress, 2017) | 0.7 | 0.9 | 90 | 7 | 128 | 8 | 3 | 1.0 | 3 | 44 |
(Savjani et al., 2018) | 1.2 | 1.2 | 170 | 10 | 142 | 8 | 3 | 0.8 | 2.4 | 67 |
(Singh et al., 2018) | 1.2 | 1.2 | 192 | 14 | 160 | 12 | 3 | 1.0 | 3 | 102 |
This study | 0.8 | 0.9 | 230 | 32 | 288 | 36 | 1 | 3.3 | 3.3 | 902 |
Motion parameters for all subjects during high-resolution (0.8 mm) spiral-out fMRI runs. Displayed are (A) the 3 translational and 3 rotational realignment parameter traces over volumes obtained by rigid-body registration, as well as (B) corresponding framewise displacement (FD) curves (Power et al., 2012). Mean FD was well below 0.2 mm for every subject (max. 0.15 mm, mean FD and standard deviation over subjects 0.09±0.04 mm), which is often considered a very rigorous criterion for censoring of motion-contaminated data (Power et al., 2015). Individual FDs of larger than 0.5 mm occurred in only 3 volumes of 2 subjects (red dashed line threshold).
Addy, N.O., Wu, H.H., Nishimura, D.G., 2012. Simple method for MR gradient system characterization and k-space trajectory estimation. Magn. Reson. Med. 68, 120–129. https://doi.org/10.1002/mrm.23217
Ahn, C.B., Kim, J.H., Cho, Z.H., 1986. High-Speed Spiral-Scan Echo Planar NMR Imaging-I. IEEE Trans. Med. Imaging 5, 2–7. https://doi.org/10.1109/TMI.1986.4307732
Amunts, K., Malikovic, A., Mohlberg, H., Schormann, T., Zilles, K., 2000. Brodmann’s Areas 17 and 18 Brought into Stereotaxic Space—Where and How Variable? NeuroImage 11, 66–84. https://doi.org/10.1006/nimg.1999.0516
Ashburner, J., Friston, K.J., 2005. Unified segmentation. NeuroImage 26, 839–851. https://doi.org/10.1016/j.neuroimage.2005.02.018
Balla, D.Z., Sanchez-Panchuelo, R.M., Wharton, S.J., Hagberg, G.E., Scheffler, K., Francis, S.T., Bowtell, R., 2014. Functional quantitative susceptibility mapping (fQSM). NeuroImage 100, 112–124. https://doi.org/10.1016/j.neuroimage.2014.06.011
Barmet, C., Tsao, J., Pruessmann, K.P., 2005. Sensitivity encoding and B0 inhomogeneity–A simultaneous reconstruction approach, in: Proceedings of the 13th Annual Meeting of ISMRM. p. 682.
Barmet, C., Zanche, N.D., Pruessmann, K.P., 2008. Spatiotemporal magnetic field monitoring for MR. Magn. Reson. Med. 60, 187–197. https://doi.org/10.1002/mrm.21603
Barth, M., Metzler, A., Klarhöfer, M., Röll, S., Moser, E., Leibfritz, D., 1999. Functional MRI of the human motor cortex using single-shot, multiple gradient-echo spiral imaging. Magn. Reson. Imaging 17, 1239–1243. https://doi.org/10.1016/S0730-725X(99)00087-9
Beatty, P.J., Nishimura, D.G., Pauly, J.M., 2005. Rapid gridding reconstruction with a minimal oversampling ratio. IEEE Trans. Med. Imaging 24, 799–808. https://doi.org/10.1109/TMI.2005.848376
Bernstein, M.A., King, K.F., Zhou, X.J. (Eds.), 2004. Chapter 17 - Advanced Pulse Sequence Techniques, in: Handbook of MRI Pulse Sequences. Academic Press, Burlington, pp. 802–954.
Bezanson, J., Edelman, A., Karpinski, S., Shah, V.B., 2017. Julia: A Fresh Approach to Numerical Computing. SIAM Rev. 59, 65–98. https://doi.org/10.1137/141000671
Bhavsar, P.S., Zwart, N.R., Pipe, J.G., 2014. Fast, variable system delay correction for spiral MRI. Magn. Reson. Med. 71, 773–782. https://doi.org/10.1002/mrm.24730
Bianciardi, M., Gelderen, P. van, Duyn, J.H., 2014. Investigation of BOLD fMRI resonance frequency shifts and quantitative susceptibility changes at 7 T. Hum. Brain Mapp. 35, 2191–2205. https://doi.org/10.1002/hbm.22320
Blazejewska, A.I., Fischl, B., Wald, L.L., Polimeni, J.R., 2019. Intracortical smoothing of small-voxel fMRI data can provide increased detection power without spatial resolution losses compared to conventional large-voxel fMRI data. NeuroImage 189, 601–614. https://doi.org/10.1016/j.neuroimage.2019.01.054
Block, K.T., Frahm, J., 2005. Spiral imaging: A critical appraisal. J. Magn. Reson. Imaging 21, 657–668. https://doi.org/10.1002/jmri.20320
Bollmann, S., Kasper, L., Pruessmann, K., Barth, M., Stephan, K.E., 2018. Interactive and flexible quality control in fMRI sequence evaluation: the uniQC toolbox, in: Proceedings of the 26th Annual Meeting of ISMRM. Presented at the ISMRM, Paris, France, p. 2842.
Bollmann, S., Kasper, L., Vannesjo, S.J., Diaconescu, A.O., Dietrich, B.E., Gross, S., Stephan, K.E., Pruessmann, K.P., 2017. Analysis and correction of field fluctuations in fMRI data using field monitoring. NeuroImage, Cleaning up the fMRI time series: Mitigating noise with advanced acquisition and correction strategies 154, 92–105. https://doi.org/10.1016/j.neuroimage.2017.01.014
Börnert, P., Aldefeld, B., Eggers, H., 2000. Reversed spiral MR imaging. Magn. Reson. Med. 44, 479–484. https://doi.org/10.1002/1522-2594(200009)44:3\<479::AID-MRM20>3.0.CO;2-2
Börnert, P., Schomberg, H., Aldefeld, B., Groen, J., 1999. Improvements in spiral MR imaging. Magn. Reson. Mater. Phys. Biol. Med. 9, 29–41. https://doi.org/10.1007/BF02634590
Breuer, F.A., Blaimer, M., Mueller, M.F., Seiberlich, N., Heidemann, R.M., Griswold, M.A., Jakob, P.M., 2006. Controlled aliasing in volumetric parallel imaging (2D CAIPIRINHA). Magn. Reson. Med. 55, 549–556. https://doi.org/10.1002/mrm.20787
Calhoun, V.D., Adalı, T., Pearlson, G.D., Zijl, P.C.M. van, Pekar, J.J., 2002. Independent component analysis of fMRI data in the complex domain. Magn. Reson. Med. 48, 180–192. https://doi.org/10.1002/mrm.10202
Campbell-Washburn, A.E., Xue, H., Lederman, R.J., Faranesh, A.Z., Hansen, M.S., 2016. Real-time distortion correction of spiral and echo planar images using the gradient system impulse response function. Magn. Reson. Med. 75, 2278–2285. https://doi.org/10.1002/mrm.25788
Cavusoglu, M., Kasper, L., Pruessmann, K.P., 2017. Single-Shot Spiral Arterial Spin Labeling MRI Enabled by Concurrent Field Monitoring, in: Proceedings of the 25th Annual Meeting of ISMRM. Presented at the ISMRM, Honolulu, Hawaii, USA, p. 1042.
Chang, Y.V., Vidorreta, M., Wang, Z., Detre, J.A., 2017. 3D-accelerated, stack-of-spirals acquisitions and reconstruction of arterial spin labeling MRI. Magn. Reson. Med. 78, 1405–1419. https://doi.org/10.1002/mrm.26549
Cheng, K., Waggoner, R.A., Tanaka, K., 2001. Human Ocular Dominance Columns as Revealed by High-Field Functional Magnetic Resonance Imaging. Neuron 32, 359–374. https://doi.org/10.1016/S0896-6273(01)00477-9
Cohen, M.S., Schmitt, F., 2012. Echo planar imaging before and after fMRI: A personal history. NeuroImage, 20 YEARS OF fMRI 62, 652–659. https://doi.org/10.1016/j.neuroimage.2012.01.038
D’Ardenne, K., Mcclure, S.M., Nystrom, L.E., Cohen, J.D., 2008. BOLD Responses Reflecting Dopaminergic Signals in the Human Ventral Tegmental Area. Science 319, 1264―1267. http://dx.doi.org/10.1126/science.1150605
de Hollander, G., Keuken, M.C., van der Zwaag, W., Forstmann, B.U., Trampel, R., 2017. Comparing functional MRI protocols for small, iron-rich basal ganglia nuclei such as the subthalamic nucleus at 7 T and 3 T. Hum. Brain Mapp. 38, 3226–3248. https://doi.org/10.1002/hbm.23586
De Martino, F., Moerel, M., van de Moortele, P.-F., Ugurbil, K., Goebel, R., Yacoub, E., Formisano, E., 2013. Spatial organization of frequency preference and selectivity in the human inferior colliculus. Nat. Commun. 4, 1386. https://doi.org/10.1038/ncomms2379
Deng, W., Zahneisen, B., Stenger, V.A., 2016. Rotated stack-of-spirals partial acquisition for rapid volumetric parallel MRI. Magn. Reson. Med. 76, 127–135. https://doi.org/10.1002/mrm.25863
Detre, J.A., Leigh, J.S., Williams, D.S., Koretsky, A.P., 1992. Perfusion imaging. Magn. Reson. Med. 23, 37–45. https://doi.org/10.1002/mrm.1910230106
Detre, J.A., Rao, H., Wang, D.J.J., Chen, Y.F., Wang, Z., 2012. Applications of arterial spin labeled MRI in the brain. J. Magn. Reson. Imaging 35, 1026–1037. https://doi.org/10.1002/jmri.23581
Dietrich, B.E., Brunner, D.O., Wilm, B.J., Barmet, C., Gross, S., Kasper, L., Haeberlin, M., Schmid, T., Vannesjo, S.J., Pruessmann, K.P., 2016a. A field camera for MR sequence monitoring and system analysis: MR Sequence Monitoring and System Analysis Camera. Magn. Reson. Med. 75, 1831–1840. https://doi.org/10.1002/mrm.25770
Dietrich, B.E., Reber, J., Brunner, D.O., Wilm, B.J., Pruessmann, K.P., 2016b. Analysis and Prediction of Gradient Response Functions under Thermal Load, in: Proceedings of the 24th Annual Meeting of ISMRM. Presented at the ISMRM, Singapore, Singapore, p. 3551.
Duerst, Y., Wilm, B.J., Wyss, M., Dietrich, B.E., Gross, S., Schmid, T., Brunner, D.O., Pruessmann, K.P., 2016. Utility of real-time field control in T2*-Weighted head MRI at 7T. Magn. Reson. Med. 76, 430–439. https://doi.org/10.1002/mrm.25838
Duyn, J.H., Yang, Y., Frank, J.A., van der Veen, J.W., 1998. Simple Correction Method fork-Space Trajectory Deviations in MRI. J. Magn. Reson. 132, 150–153. https://doi.org/10.1006/jmre.1998.1396
Eickhoff, S.B., Heim, S., Zilles, K., Amunts, K., 2006. Testing anatomically specified hypotheses in functional imaging using cytoarchitectonic maps. NeuroImage 32, 570–582. https://doi.org/10.1016/j.neuroimage.2006.04.204
Eickhoff, S.B., Paus, T., Caspers, S., Grosbras, M.-H., Evans, A.C., Zilles, K., Amunts, K., 2007. Assignment of functional activations to probabilistic cytoarchitectonic areas revisited. NeuroImage 36, 511–521. https://doi.org/10.1016/j.neuroimage.2007.03.060
Eickhoff, S.B., Stephan, K.E., Mohlberg, H., Grefkes, C., Fink, G.R., Amunts, K., Zilles, K., 2005. A new SPM toolbox for combining probabilistic cytoarchitectonic maps and functional imaging data. NeuroImage 25, 1325–1335. https://doi.org/10.1016/j.neuroimage.2004.12.034
Engel, M., Kasper, L., Barmet, C., Schmid, T., Vionnet, L., Wilm, B., Pruessmann, K.P., 2018. Single-shot spiral imaging at 7 T. Magn. Reson. Med. 80, 1836–1846. https://doi.org/10.1002/mrm.27176
Engel, M., Kasper, L., Wilm, B., Dietrich, B., Vionnet, L., Hennel, F., Reber, J., Pruessmann, K.P., 2021. T-Hex: Tilted hexagonal grids for rapid 3D imaging. Magn. Reson. Med. 85, 2507–2523. https://doi.org/10.1002/mrm.28600
Engel, S.A., Glover, G.H., Wandell, B.A., 1997. Retinotopic organization in human visual cortex and the spatial precision of functional MRI. Cereb. Cortex 7, 181–192. https://doi.org/10.1093/cercor/7.2.181
Feinberg, D.A., Vu, A.T., Beckett, A., 2018. Pushing the limits of ultra-high resolution human brain imaging with SMS-EPI demonstrated for columnar level fMRI. NeuroImage, Pushing the spatio-temporal limits of MRI and fMRI 164, 155–163. https://doi.org/10.1016/j.neuroimage.2017.02.020
Fessler, J.A., 2010. Model-Based Image Reconstruction for MRI. IEEE Signal Process. Mag. 27, 81–89. https://doi.org/10.1109/MSP.2010.936726
Foo, T.K.F., Laskaris, E., Vermilyea, M., Xu, M., Thompson, P., Conte, G., Epps, C.V., Immer, C., Lee, S.-K., Tan, E.T., Graziani, D., Mathieu, J.-B., Hardy, C.J., Schenck, J.F., Fiveland, E., Stautner, W., Ricci, J., Piel, J., Park, K., Hua, Y., Bai, Y., Kagan, A., Stanley, D., Weavers, P.T., Gray, E., Shu, Y., Frick, M.A., Campeau, N.G., Trzasko, J., Huston, J., Bernstein, M.A., 2018. Lightweight, compact, and high-performance 3T MR system for imaging the brain and extremities. Magn. Reson. Med. 80, 2232–2245. https://doi.org/10.1002/mrm.27175
Fracasso, A., Petridou, N., Dumoulin, S.O., 2016. Systematic variation of population receptive field properties across cortical depth in human visual cortex. NeuroImage 139, 427–438. https://doi.org/10.1016/j.neuroimage.2016.06.048
Frässle, S., Aponte, E.A., Bollmann, S., Brodersen, K.H., Do, C.T., Harrison, O.K., Harrison, S.J., Heinzle, J., Iglesias, S., Kasper, L., Lomakina, E.I., Mathys, C., Müller-Schrader, M., Pereira, I., Petzschner, F.H., Raman, S., Schöbi, D., Toussaint, B., Weber, L.A., Yao, Y., Stephan, K.E., 2021. TAPAS: An Open-Source Software Package for Translational Neuromodeling and Computational Psychiatry. Front. Psychiatry 12. https://doi.org/10.3389/fpsyt.2021.680811
Friston, K.J., 2007. Chapter 2 - Statistical parametric mapping, in: Friston, K.J., Ashburner, J., Kiebel, S., Nichols, T., Penny, W. (Eds.), Statistical Parametric Mapping. Academic Press, London, pp. 10–31.
Friston, K.J., Williams, S., Howard, R., Frackowiak, R.S.J., Turner, R., 1996. Movement-Related effects in fMRI time-series. Magn. Reson. Med. 35, 346–355. https://doi.org/10.1002/mrm.1910350312
Glover, G.H., 2012. Spiral imaging in fMRI. NeuroImage, 20 Years of fMRI 62, 706–712. https://doi.org/10.1016/j.neuroimage.2011.10.039
Glover, G.H., 1999. Simple analytic spiral K-space algorithm. Magn. Reson. Med. 42, 412–415. https://doi.org/10.1002/(SICI)1522-2594(199908)42:2\<412::AID-MRM25>3.0.CO;2-U
Glover, G.H., Law, C.S., 2001. Spiral-in/out BOLD fMRI for increased SNR and reduced susceptibility artifacts. Magn. Reson. Med. 46, 515–522. https://doi.org/10.1002/mrm.1222
Glover, G.H., Li, T.-Q., Ress, D., 2000. Image-based method for retrospective correction of physiological motion effects in fMRI: RETROICOR. Magn. Reson. Med. 44, 162–167. https://doi.org/10.1002/1522-2594(200007)44:1\<162::AID-MRM23>3.0.CO;2-E
Glover, G.H., Thomason, M.E., 2004. Improved combination of spiral-in/out images for BOLD fMRI. Magn. Reson. Med. 51, 863–868. https://doi.org/10.1002/mrm.20016
Gorgolewski, K.J., Varoquaux, G., Rivera, G., Schwarz, Y., Ghosh, S.S., Maumet, C., Sochat, V.V., Nichols, T.E., Poldrack, R.A., Poline, J.-B., Yarkoni, T., Margulies, D.S., 2015. NeuroVault.org: a web-based repository for collecting and sharing unthresholded statistical maps of the human brain. Front. Neuroinformatics 9. https://doi.org/10.3389/fninf.2015.00008
Graedel, N.N., Hurley, S.A., Clare, S., Miller, K.L., Pruessmann, K.P., Vannesjo, S.J., 2017. Comparison of gradient impulse response functions measured with a dynamic field camera and a phantom-based technique. Presented at the 34th ESMRMB Congress, Barcelona/ES, p. 378.
Griswold, M.A., Jakob, P.M., Heidemann, R.M., Nittka, M., Jellus, V., Wang, J., Kiefer, B., Haase, A., 2002. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn. Reson. Med. 47, 1202–1210. https://doi.org/10.1002/mrm.10171
Harvey, A.K., Pattinson, K.T.S., Brooks, J.C.W., Mayhew, S.D., Jenkinson, M., Wise, R.G., 2008. Brainstem functional magnetic resonance imaging: Disentangling signal from physiological noise. J. Magn. Reson. Imaging 28, 1337–1344. https://doi.org/10.1002/jmri.21623
Heidemann, R.M., Griswold, M.A., Seiberlich, N., Krüger, G., Kannengiesser, S.A.R., Kiefer, B., Wiggins, G., Wald, L.L., Jakob, P.M., 2006. Direct parallel image reconstructions for spiral trajectories using GRAPPA. Magn. Reson. Med. 56, 317–326. https://doi.org/10.1002/mrm.20951
Hernando, D., Haldar, J.P., Sutton, B.P., Ma, J., Kellman, P., Liang, Z.-P., 2008. Joint estimation of water/fat images and field inhomogeneity map. Magn. Reson. Med. 59, 571–580. https://doi.org/10.1002/mrm.21522
Huber, L., Handwerker, D.A., Jangraw, D.C., Chen, G., Hall, A., Stüber, C., Gonzalez-Castillo, J., Ivanov, D., Marrett, S., Guidi, M., Goense, J., Poser, B.A., Bandettini, P.A., 2017a. High-Resolution CBV-fMRI Allows Mapping of Laminar Activity and Connectivity of Cortical Input and Output in Human M1. Neuron 96, 1253-1263.e7. https://doi.org/10.1016/j.neuron.2017.11.005
Huber, L., Ivanov, D., Handwerker, D.A., Marrett, S., Guidi, M., Uludağ, K., Bandettini, P.A., Poser, B.A., 2018. Techniques for blood volume fMRI with VASO: From low-resolution mapping towards sub-millimeter layer-dependent applications. NeuroImage 164, 131–143. https://doi.org/10.1016/j.neuroimage.2016.11.039
Huber, L., Uludağ, K., Möller, H.E., 2017b. Non-BOLD contrast for laminar fMRI in humans: CBF, CBV, and CMRO2. NeuroImage. https://doi.org/10.1016/j.neuroimage.2017.07.041
Jackson, J.I., Meyer, C.H., Nishimura, D.G., Macovski, A., 1991. Selection of a convolution function for Fourier inversion using gridding. IEEE Trans. Med. Imaging 10, 473―478. http://dx.doi.org/10.1109/42.97598
Jung, Y., Samsonov, A.A., Liu, T.T., Buracas, G.T., 2013. High efficiency multishot interleaved spiral-in/out: Acquisition for high-resolution BOLD fMRI. Magn. Reson. Med. 70, 420–428. https://doi.org/10.1002/mrm.24476
Kaldoudi, E., Williams, S.C.R., Barker, G.J., Tofts, P.S., 1993. A chemical shift selective inversion recovery sequence for fat-suppressed MRI: Theory and experimental validation. Magn. Reson. Imaging 11, 341–355. https://doi.org/10.1016/0730-725X(93)90067-N
Kashyap, S., Ivanov, D., Havlicek, M., Sengupta, S., Poser, B.A., Uludağ, K., 2018. Resolving laminar activation in human V1 using ultra-high spatial resolution fMRI at 7T. Sci. Rep. 8, 1–11. https://doi.org/10.1038/s41598-018-35333-3
Kasper, L., Bollmann, S., Diaconescu, A.O., Hutton, C., Heinzle, J., Iglesias, S., Hauser, T.U., Sebold, M., Manjaly, Z.-M., Pruessmann, K.P., Stephan, K.E., 2017. The PhysIO Toolbox for Modeling Physiological Noise in fMRI Data. J. Neurosci. Methods 276, 56–72. https://doi.org/10.1016/j.jneumeth.2016.10.019
Kasper, L., Bollmann, S., Vannesjo, S.J., Gross, S., Haeberlin, M., Dietrich, B.E., Pruessmann, K.P., 2015. Monitoring, analysis, and correction of magnetic field fluctuations in echo planar imaging time series. Magn. Reson. Med. 74, 396–409. https://doi.org/10.1002/mrm.25407
Kasper, L., Engel, M., Barmet, C., Haeberlin, M., Wilm, B.J., Dietrich, B.E., Schmid, T., Gross, S., Brunner, D.O., Stephan, K.E., Pruessmann, K.P., 2018. Rapid anatomical brain imaging using spiral acquisition and an expanded signal model. NeuroImage, Neuroimaging with Ultra-high Field MRI: Present and Future 168, 88–100. https://doi.org/10.1016/j.neuroimage.2017.07.062
Kasper, L., Engel, M., Heinzle, J., Mueller-Schrader, M., Graedel, N.N., Reber, J., Schmid, T., Barmet, C., Wilm, B.J., Stephan, K.E., Pruessmann, K.P., 2021. Data Collection for the Article “Advances in Spiral fMRI - A High-resolution Study with Single-shot Acquisition”: Single Subject Dataset. https://doi.org/10.3929/ethz-b-000487412
Kasper, L., Haeberlin, M., Dietrich, B.E., Gross, S., Barmet, C., Wilm, B.J., Vannesjo, S.J., Brunner, D.O., Ruff, C.C., Stephan, K.E., Pruessmann, K.P., 2014. Matched-filter acquisition for BOLD fMRI. NeuroImage 100, 145–160. https://doi.org/10.1016/j.neuroimage.2014.05.024
Katyal, S., Zughni, S., Greene, C., Ress, D., 2010. Topography of Covert Visual Attention in Human Superior Colliculus. J. Neurophysiol. 104, 3074–3083. https://doi.org/10.1152/jn.00283.2010
Keeling, S.L., Bammer, R., 2004. A variational approach to magnetic resonance coil sensitivity estimation. Appl. Math. Comput. 158, 359–388. https://doi.org/10.1016/j.amc.2003.08.110
Kimmlingen, R., Eberlein, E., Dietz, P., Kreher, S., Schuster, J., Riegler, J., Matschl, V., Schnetter, V., Schmidt, A., Lenz, H., Mustafa, E., Fischer, D., Potthast, A., Kreischer, L., Eberler, M., Hebrank, F., Thein, H., Heberlein, K., Hoecht, P., Witzel, T., Tisdall, D., Xu, J., Yacoub, E., Adriany, G., Auerbach, E., Moeller, S., Feinberg, D., Lehne, D., Wald, L.L., Rosen, B., Ugurbil, K., Essen, D. van, Wedeen, V., Schmitt, F., 2012. Concept and realization of high strength gradients for the Human Connectome Project, in: Proceedings of the 20th Annual Meeting of ISMRM. Presented at the ISMRM, Melbourne, Australia, p. 696.
Knopp, T., Grosser, M., 2021. MRIReco.jl: An MRI reconstruction framework written in Julia. Magn. Reson. Med. 86, 1633–1646. https://doi.org/10.1002/mrm.28792
Kok, P., Bains, L.J., van Mourik, T., Norris, D.G., de Lange, F.P., 2016. Selective Activation of the Deep Layers of the Human Primary Visual Cortex by Top-Down Feedback. Curr. Biol. 26, 371–376. https://doi.org/10.1016/j.cub.2015.12.038
Koopmans, P.J., Barth, M., Norris, D.G., 2010. Layer-specific BOLD activation in human V1. Hum. Brain Mapp. 31, 1297–1304. https://doi.org/10.1002/hbm.20936
Kujovic, M., Zilles, K., Malikovic, A., Schleicher, A., Mohlberg, H., Rottschy, C., Eickhoff, S.B., Amunts, K., 2013. Cytoarchitectonic mapping of the human dorsal extrastriate cortex. Brain Struct. Funct. 218, 157–172. https://doi.org/10.1007/s00429-012-0390-9
Kundu, P., Inati, S.J., Evans, J.W., Luh, W.-M., Bandettini, P.A., 2012. Differentiating BOLD and non-BOLD signals in fMRI time series using multi-echo EPI. NeuroImage 60, 1759–1770. https://doi.org/10.1016/j.neuroimage.2011.12.028
Kurban, D., Liberman, G., Kashyap, S., Ivanov, D., Poser, B.A., 2019. Dual-echo simultaneous multi-slice spiral acquisition for simultaneous CBF and BOLD fMRI at 7T, in: Proceedings of the 27th Annual Meeting of ISMRM. Montreal, Canada, p. 367.
Larkman, D.J., 2007. The g-Factor and Coil Design, in: Schoenberg, S.O., Dietrich, O., Reiser, M.F. (Eds.), Parallel Imaging in Clinical MR Applications, Medical Radiology. Springer, Berlin, Heidelberg, pp. 37–48. https://doi.org/10.1007/978-3-540-68879-2_3
Larkman, D.J., Hajnal, J.V., Herlihy, A.H., Coutts, G.A., Young, I.R., Ehnholm, G., 2001. Use of multicoil arrays for separation of signal from multiple slices simultaneously excited. J. Magn. Reson. Imaging 13, 313–317. https://doi.org/10.1002/1522-2586(200102)13:2\<313::AID-JMRI1045>3.0.CO;2-W
Law, C.S., Glover, G.H., 2009. Interleaved spiral-in/out with application to functional MRI (fMRI). Magn. Reson. Med. 62, 829–834. https://doi.org/10.1002/mrm.22056
Lawrence, S.J.D., van Mourik, T., Kok, P., Koopmans, P.J., Norris, D.G., de Lange, F.P., 2018. Laminar Organization of Working Memory Signals in Human Visual Cortex. Curr. Biol. 28, 3435-3440.e4. https://doi.org/10.1016/j.cub.2018.08.043
Lee, Y., Wilm, B.J., Brunner, D.O., Gross, S., Schmid, T., Nagy, Z., Pruessmann, K.P., 2021. On the signal-to-noise ratio benefit of spiral acquisition in diffusion MRI. Magn. Reson. Med. 85, 1924–1937. https://doi.org/10.1002/mrm.28554
Lewis, L.D., Setsompop, K., Rosen, B.R., Polimeni, J.R., 2016. Fast fMRI can detect oscillatory neural activity in humans. Proc. Natl. Acad. Sci. 113, E6679–E6685. https://doi.org/10.1073/pnas.1608117113
Likes, R.S., 1981. Moving gradient zeugmatography. US 4307343.
Lu, H., Golay, X., Pekar, J.J., Zijl, P.C.M. van, 2003. Functional magnetic resonance imaging based on changes in vascular space occupancy. Magn. Reson. Med. 50, 263–274. https://doi.org/10.1002/mrm.10519
Lu, H., Hua, J., van Zijl, P.C.M., 2013. Noninvasive functional imaging of cerebral blood volume with vascular-space-occupancy (VASO) MRI. NMR Biomed. 26, 932–948. https://doi.org/10.1002/nbm.2905
Lustig, M., Kim, S.-J., Pauly, J.M., 2008. A Fast Method for Designing Time-Optimal Gradient Waveforms for Arbitrary k-Space Trajectories. IEEE Trans. Med. Imaging 27, 866–873. https://doi.org/10.1109/TMI.2008.922699
Lustig, M., Pauly, J.M., 2010. SPIRiT: Iterative self-consistent parallel imaging reconstruction from arbitrary k-space. Magn. Reson. Med. 64, 457–471. https://doi.org/10.1002/mrm.22428
Maier, O., Baete, S.H., Fyrdahl, A., Hammernik, K., Harrevelt, S., Kasper, L., Karakuzu, A., Loecher, M., Patzig, F., Tian, Y., Wang, K., Gallichan, D., Uecker, M., Knoll, F., 2021. CG-SENSE revisited: Results from the first ISMRM reproducibility challenge. Magn. Reson. Med. 85, 1821–1839. https://doi.org/10.1002/mrm.28569
Malikovic, A., Amunts, K., Schleicher, A., Mohlberg, H., Eickhoff, S.B., Wilms, M., Palomero-Gallagher, N., Armstrong, E., Zilles, K., 2007. Cytoarchitectonic Analysis of the Human Extrastriate Cortex in the Region of V5/MT+: A Probabilistic, Stereotaxic Map of Area hOc5. Cereb. Cortex 17, 562–574. https://doi.org/10.1093/cercor/bhj181
Malikovic, A., Amunts, K., Schleicher, A., Mohlberg, H., Kujovic, M., Palomero-Gallagher, N., Eickhoff, S.B., Zilles, K., Zilles, K., 2016. Cytoarchitecture of the human lateral occipital cortex: mapping of two extrastriate areas hOc4la and hOc4lp. Brain Struct. Funct. 221, 1877–1897. https://doi.org/10.1007/s00429-015-1009-8
Man, L.-C., Pauly, J.M., Macovski, A., 1997. Multifrequency interpolation for fast off-resonance correction. Magn. Reson. Med. 37, 785–792. https://doi.org/10.1002/mrm.1910370523
Mansfield, P., 1977. Multi-planar image formation using NMR spin echoes. J. Phys. C Solid State Phys. 10, L55. https://doi.org/10.1088/0022-3719/10/3/004
Martino, F.D., Moerel, M., Ugurbil, K., Goebel, R., Yacoub, E., Formisano, E., 2015. Frequency preference and attention effects across cortical depths in the human primary auditory cortex. Proc. Natl. Acad. Sci. 112, 16036–16041. https://doi.org/10.1073/pnas.1507552112
Menon, R.S., 2002. Postacquisition suppression of large-vessel BOLD signals in high-resolution fMRI. Magn. Reson. Med. 47, 1–9. https://doi.org/10.1002/mrm.10041
Meyer, C.H., Hu, B.S., Nishimura, D.G., Macovski, A., 1992. Fast Spiral Coronary Artery Imaging. Magn. Reson. Med. 28, 202–213. https://doi.org/10.1002/mrm.1910280204
Muckli, L., De Martino, F., Vizioli, L., Petro, L.S., Smith, F.W., Ugurbil, K., Goebel, R., Yacoub, E., 2015. Contextual Feedback to Superficial Layers of V1. Curr. Biol. 25, 2690–2695. https://doi.org/10.1016/j.cub.2015.08.057
Noll, D.C., Cohen, J.D., Meyer, C.H., Schneider, W., 1995. Spiral K-space MR imaging of cortical activation. J. Magn. Reson. Imaging 5, 49–56. https://doi.org/10.1002/jmri.1880050112
Nussbaum, J., Rösler, M.B., Dietrich, B.E., Pruessmann, K.P., 2019. Nonlinearity and thermal effects in gradient chains: a cascade analysis based on current and field sensing, in: Proceedings of the 27th Annual Meeting of ISMRM. Montreal, Canada, p. 213.
Patzig, F., Kasper, L., Ulrich, T., Engel, M., Vannesjo, S.J., Weiger, M., Brunner, D.O., Wilm, B.J., Pruessmann, K.P., 2019. ISMRM 2019 RRSG Challenge: MRI Technology and Methods Lab, ETH Zurich, GitHub Submission, in: ISMRM 2019 Reproducible Research Study Group Challenge. https://dx.doi.org/10.24433/CO.5840424.v1
Patzig, F., Wilm, B.J., Gross, S., Brunner, D.O., Pruessmann, K.P., 2020. Off-Resonance Self-Correction for Single-Shot Imaging, in: Proceedings of the 28th Annual Meeting of ISMRM. p. 3040.
Peters, A.M., Brookes, M.J., Hoogenraad, F.G., Gowland, P.A., Francis, S.T., Morris, P.G., Bowtell, R., 2007. T2* measurements in human brain at 1.5, 3 and 7 T. Magn. Reson. Imaging, Proceedings of the International School on Magnetic Resonance and Brain Function 25, 748–753. https://doi.org/10.1016/j.mri.2007.02.014
Pfeuffer, J., Van de Moortele, P.-F., Ugurbil, K., Hu, X., Glover, G.H., 2002. Correction of physiologically induced global off-resonance effects in dynamic echo-planar and spiral functional imaging. Magn. Reson. Med. 47, 344–353. https://doi.org/10.1002/mrm.10065
Poser, B.A., Koopmans, P.J., Witzel, T., Wald, L.L., Barth, M., 2010. Three dimensional echo-planar imaging at 7 Tesla. NeuroImage 51, 261–266. https://doi.org/10.1016/j.neuroimage.2010.01.108
Power, J.D., Barnes, K.A., Snyder, A.Z., Schlaggar, B.L., Petersen, S.E., 2012. Spurious but systematic correlations in functional connectivity MRI networks arise from subject motion. NeuroImage 59, 2142–2154. https://doi.org/10.1016/j.neuroimage.2011.10.018
Power, J.D., Lynch, C.J., Silver, B.M., Dubin, M.J., Martin, A., Jones, R.M., 2019. Distinctions among real and apparent respiratory motions in human fMRI data. NeuroImage 201, 116041. https://doi.org/10.1016/j.neuroimage.2019.116041
Power, J.D., Plitt, M., Gotts, S.J., Kundu, P., Voon, V., Bandettini, P.A., Martin, A., 2018. Ridding fMRI data of motion-related influences: Removal of signals with distinct spatial and physical bases in multiecho data. Proc. Natl. Acad. Sci. 201720985. https://doi.org/10.1073/pnas.1720985115
Power, J.D., Schlaggar, B.L., Petersen, S.E., 2015. Recent progress and outstanding issues in motion correction in resting state fMRI. NeuroImage 105, 536–551. https://doi.org/10.1016/j.neuroimage.2014.10.044
Preston, A.R., Bornstein, A.M., Hutchinson, J.B., Gaare, M.E., Glover, G.H., Wagner, A.D., 2010. High-resolution fMRI of Content-sensitive Subsequent Memory Responses in Human Medial Temporal Lobe. J. Cogn. Neurosci. 22, 156–173. https://doi.org/10.1162/jocn.2009.21195
Pruessmann, K.P., 2006. Encoding and reconstruction in parallel MRI. NMR Biomed. 19, 288. http://dx.doi.org/10.1002/nbm.1042
Pruessmann, K.P., Weiger, M., Börnert, P., Boesiger, P., 2001. Advances in sensitivity encoding with arbitrary k-space trajectories. Magn. Reson. Med. 46, 638―651. http://dx.doi.org/10.1002/mrm.1241
Pruessmann, K.P., Weiger, M., Scheidegger, M.B., Boesiger, P., 1999. SENSE: Sensitivity encoding for fast MRI. Magn. Reson. Med. 42, 952–962. https://doi.org/10.1002/(SICI)1522-2594(199911)42:5\<952::AID-MRM16>3.0.CO;2-S
Qin, Q., 2012. Point spread functions of the T2 decay in k-space trajectories with long echo train. Magn. Reson. Imaging 30, 1134–1142. https://doi.org/10.1016/j.mri.2012.04.017
Rahmer, J., Mazurkewitz, P., Börnert, P., Nielsen, T., 2019. Rapid acquisition of the 3D MRI gradient impulse response function using a simple phantom measurement. Magn. Reson. Med. 82, 2146–2159. https://doi.org/10.1002/mrm.27902
Rahmer, J., Schmale, I., Mazurkewitz, P., Lips, O., Börnert, P., 2021. Non-Cartesian k-space trajectory calculation based on concurrent reading of the gradient amplifiers’ output currents. Magn. Reson. Med. 85, 3060–3070. https://doi.org/10.1002/mrm.28725
Ress, D., Glover, G.H., Liu, J., Wandell, B., 2007. Laminar profiles of functional activity in the human brain. NeuroImage 34, 74–84. https://doi.org/10.1016/j.neuroimage.2006.08.020
Robison, R.K., Devaraj, A., Pipe, J.G., 2010. Fast, simple gradient delay estimation for spiral MRI. Magn. Reson. Med. 63, 1683–1690. https://doi.org/10.1002/mrm.22327
Robison, R.K., Li, Z., Wang, D., Ooi, M.B., Pipe, J.G., 2019. Correction of B0 eddy current effects in spiral MRI. Magn. Reson. Med. 0. https://doi.org/10.1002/mrm.27583
Rottschy, C., Eickhoff, S.B., Schleicher, A., Mohlberg, H., Kujovic, M., Zilles, K., Amunts, K., 2007. Ventral visual cortex in humans: Cytoarchitectonic mapping of two extrastriate areas. Hum. Brain Mapp. 28, 1045–1059. https://doi.org/10.1002/hbm.20348
Savjani, R.R., Katyal, S., Halfen, E., Kim, J.H., Ress, D., 2018. Polar-angle representation of saccadic eye movements in human superior colliculus. NeuroImage 171, 199–208. https://doi.org/10.1016/j.neuroimage.2017.12.080
Schmitt, F., Stehling, M.K., Turner, R., 2012. Echo-Planar Imaging: Theory, Technique and Application. Springer Science & Business Media.
Setsompop, K., Gagoski, B.A., Polimeni, J.R., Witzel, T., Wedeen, V.J., Wald, L.L., 2012. Blipped-controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging with reduced g-factor penalty. Magn. Reson. Med. 67, 1210–1224. https://doi.org/10.1002/mrm.23097
Shewchuk, J.R., 1994. An Introduction to the Conjugate Gradient Method Without the Agonizing Pain.
Siero, J.C., Petridou, N., Hoogduin, H., Luijten, P.R., Ramsey, N.F., 2011. Cortical Depth-Dependent Temporal Dynamics of the BOLD Response in the Human Brain: J. Cereb. Blood Flow Metab. https://doi.org/10.1038/jcbfm.2011.57
Singh, V., Pfeuffer, J., Zhao, T., Ress, D., 2018. Evaluation of spiral acquisition variants for functional imaging of human superior colliculus at 3T field strength. Magn. Reson. Med. 79, 1931–1940. https://doi.org/10.1002/mrm.26845
Sitek, K.R., Gulban, O.F., Calabrese, E., Johnson, G.A., Lage-Castellanos, A., Moerel, M., Ghosh, S.S., De Martino, F., 2019. Mapping the human subcortical auditory system using histology, postmortem MRI and in vivo MRI at 7T. eLife 8, e48932. https://doi.org/10.7554/eLife.48932
Smith, S.M., Beckmann, C.F., Andersson, J., Auerbach, E.J., Bijsterbosch, J., Douaud, G., Duff, E., Feinberg, D.A., Griffanti, L., Harms, M.P., Kelly, M., Laumann, T., Miller, K.L., Moeller, S., Petersen, S., Power, J., Salimi-Khorshidi, G., Snyder, A.Z., Vu, A.T., Woolrich, M.W., Xu, J., Yacoub, E., Uğurbil, K., Van Essen, D.C., Glasser, M.F., 2013. Resting-state fMRI in the Human Connectome Project. NeuroImage, Mapping the Connectome 80, 144–168. https://doi.org/10.1016/j.neuroimage.2013.05.039
Sodickson, D.K., Manning, W.J., 1997. Simultaneous acquisition of spatial harmonics (SMASH): Fast imaging with radiofrequency coil arrays. Magn. Reson. Med. 38, 591–603. https://doi.org/10.1002/mrm.1910380414
Splitthoff, D. n., Zaitsev, M., 2009. SENSE shimming (SSH): A fast approach for determining B0 field inhomogeneities using sensitivity coding. Magn. Reson. Med. 62, 1319–1325. https://doi.org/10.1002/mrm.22083
Stich, M., Pfaff, C., Wech, T., Slawig, A., Ruyters, G., Dewdney, A., Ringler, R., Köstler, H., 2020. The temperature dependence of gradient system response characteristics. Magn. Reson. Med. 83, 1519–1527. https://doi.org/10.1002/mrm.28013
Stikov, N., Trzasko, J.D., Bernstein, M.A., 2019. Reproducibility and the future of MRI research. Magn. Reson. Med. 82, 1981–1983. https://doi.org/10.1002/mrm.27939
Sutton, B.P., Noll, D.C., Fessler, J.A., 2003. Fast, iterative image reconstruction for MRI in the presence of field inhomogeneities. IEEE Trans. Med. Imaging 22, 178–188. https://doi.org/10.1109/TMI.2002.808360
Uğurbil, K., Xu, J., Auerbach, E.J., Moeller, S., Vu, A.T., Duarte-Carvajalino, J.M., Lenglet, C., Wu, X., Schmitter, S., Van de Moortele, P.F., Strupp, J., Sapiro, G., De Martino, F., Wang, D., Harel, N., Garwood, M., Chen, L., Feinberg, D.A., Smith, S.M., Miller, K.L., Sotiropoulos, S.N., Jbabdi, S., Andersson, J.L.R., Behrens, T.E.J., Glasser, M.F., Van Essen, D.C., Yacoub, E., 2013. Pushing spatial and temporal resolution for functional and diffusion MRI in the Human Connectome Project. NeuroImage 80, 80–104. https://doi.org/10.1016/j.neuroimage.2013.05.012
Uludağ, K., Blinder, P., 2018. Linking brain vascular physiology to hemodynamic response in ultra-high field MRI. NeuroImage 168, 279–295. https://doi.org/10.1016/j.neuroimage.2017.02.063
Vannesjo, S.J., Dietrich, B.E., Pavan, M., Brunner, D.O., Wilm, B.J., Barmet, C., Pruessmann, K.P., 2014. Field camera measurements of gradient and shim impulse responses using frequency sweeps. Magn. Reson. Med. 72, 570–583. https://doi.org/10.1002/mrm.24934
Vannesjo, S.J., Graedel, N.N., Kasper, L., Gross, S., Busch, J., Haeberlin, M., Barmet, C., Pruessmann, K.P., 2016. Image reconstruction using a gradient impulse response model for trajectory prediction. Magn. Reson. Med. 76, 45–58. https://doi.org/10.1002/mrm.25841
Vannesjo, S.J., Haeberlin, M., Kasper, L., Pavan, M., Wilm, B.J., Barmet, C., Pruessmann, K.P., 2013. Gradient system characterization by impulse response measurements with a dynamic field camera. Magn. Reson. Med. 69, 583–93. https://doi.org/10.1002/mrm.24263
Wandell, B.A., Dumoulin, S.O., Brewer, A.A., 2007. Visual Field Maps in Human Cortex. Neuron 56, 366–383. https://doi.org/10.1016/j.neuron.2007.10.012
Wang, Y.C., Bianciardi, M., Chanes, L., Satpute, A.B., 2020. Ultra High Field fMRI of Human Superior Colliculi Activity during Affective Visual Processing. Sci. Rep. 10, 1331. https://doi.org/10.1038/s41598-020-57653-z
Warnking, J., Dojat, M., Guérin-Dugué, A., Delon-Martin, C., Olympieff, S., Richard, N., Chéhikian, A., Segebarth, C., 2002. fMRI Retinotopic Mapping—Step by Step. NeuroImage 17, 1665–1683. https://doi.org/10.1006/nimg.2002.1304
Weiger, M., Overweg, J., Rösler, M.B., Froidevaux, R., Hennel, F., Wilm, B.J., Penn, A., Sturzenegger, U., Schuth, W., Mathlener, M., Borgo, M., Börnert, P., Leussler, C., Luechinger, R., Dietrich, B.E., Reber, J., Brunner, D.O., Schmid, T., Vionnet, L., Pruessmann, K.P., 2018. A high-performance gradient insert for rapid and short-T2 imaging at full duty cycle. Magn. Reson. Med. 79, 3256–3266. https://doi.org/10.1002/mrm.26954
Weiger, M., Pruessmann, K.P., Österbauer, R., Börnert, P., Boesiger, P., Jezzard, P., 2002. Sensitivity-encoded single-shot spiral imaging for reduced susceptibility artifacts in BOLD fMRI. Magn. Reson. Med. 48, 860–866. https://doi.org/10.1002/mrm.10286
Welvaert, M., Rosseel, Y., 2013. On the Definition of Signal-To-Noise Ratio and Contrast-To-Noise Ratio for fMRI Data. PLoS ONE 8, e77089. https://doi.org/10.1371/journal.pone.0077089
Wilkinson, M.D., Dumontier, M., Aalbersberg, Ij.J., Appleton, G., Axton, M., Baak, A., Blomberg, N., Boiten, J.-W., Santos, L.B. da S., Bourne, P.E., Bouwman, J., Brookes, A.J., Clark, T., Crosas, M., Dillo, I., Dumon, O., Edmunds, S., Evelo, C.T., Finkers, R., Gonzalez-Beltran, A., Gray, A.J.G., Groth, P., Goble, C., Grethe, J.S., Heringa, J., Hoen, P.A.C. ’t, Hooft, R., Kuhn, T., Kok, R., Kok, J., Lusher, S.J., Martone, M.E., Mons, A., Packer, A.L., Persson, B., Rocca-Serra, P., Roos, M., Schaik, R. van, Sansone, S.-A., Schultes, E., Sengstag, T., Slater, T., Strawn, G., Swertz, M.A., Thompson, M., Lei, J. van der, Mulligen, E. van, Velterop, J., Waagmeester, A., Wittenburg, P., Wolstencroft, K., Zhao, J., Mons, B., 2016. The FAIR Guiding Principles for scientific data management and stewardship. Sci. Data 3, 1–9. https://doi.org/10.1038/sdata.2016.18
Wilm, B.J., Barmet, C., Gross, S., Kasper, L., Vannesjo, S.J., Haeberlin, M., Dietrich, B.E., Brunner, D.O., Schmid, T., Pruessmann, K.P., 2017. Single-shot spiral imaging enabled by an expanded encoding model: Demonstration in diffusion MRI. Magn. Reson. Med. 77, 83–91. https://doi.org/10.1002/mrm.26493
Wilm, B.J., Barmet, C., Pavan, M., Pruessmann, K.P., 2011. Higher order reconstruction for MRI in the presence of spatiotemporal field perturbations. Magn. Reson. Med. 65, 1690–1701. https://doi.org/10.1002/mrm.22767
Wright, K.L., Hamilton, J.I., Griswold, M.A., Gulani, V., Seiberlich, N., 2014. Non-Cartesian parallel imaging reconstruction. J. Magn. Reson. Imaging 40, 1022–1040. https://doi.org/10.1002/jmri.24521
Yacoub, E., Harel, N., Uğurbil, K., 2008. High-field fMRI unveils orientation columns in humans. Proc. Natl. Acad. Sci. 105, 10607–10612. https://doi.org/10.1073/pnas.0804110105
Zahneisen, B., Ernst, T., Poser, B.A., 2015. SENSE and simultaneous multislice imaging. Magn. Reson. Med. 74, 1356–1362. https://doi.org/10.1002/mrm.25519
Zahneisen, B., Poser, B.A., Ernst, T., Stenger, A.V., 2014. Simultaneous Multi-Slice fMRI using spiral trajectories. NeuroImage 92, 8–18. https://doi.org/10.1016/j.neuroimage.2014.01.056
Zanche, N.D., Barmet, C., Nordmeyer‐Massner, J.A., Pruessmann, K.P., 2008. NMR probes for measuring magnetic fields and field dynamics in MR systems. Magn. Reson. Med. 60, 176–186. https://doi.org/10.1002/mrm.21624